Clin Biochem. 2016 Apr;49(6):502-504. doi: 10.1016/j.clinbiochem.2015.11.015. Epub 2015 Nov 21.
Since HbA1c became a diagnostic criterion for type 2 diabetes, assays measuring this analyte were required to be analytically stronger. Although precision of Sebia Capillarys 2 Flex Piercing® was previously evaluated with standard methods and material, we wanted to document the long-term performance of this analyzer in two different centers.
In this study, we evaluated, with a non-conventional methodology, the analytical precision of the Sebia Capillarys 2 Flex Piercing® using HbA1c patient data and compared it with two different common commercial immunoassays.
CV of 1.33% and 1.32% was obtained in the two centers for the electrophoretic technique compared to 2.66% and 3.92% for the Roche Cobas Integra 800® and Beckman Synchron DxC® immunoassays technique respectively.
Our results show that capillary electrophoresis is more robust than the 2 immunoassays tested for long-term precision and even exceeds minimal requirements recommended.