Suppr超能文献

无酶化学发光免疫分析法测定促甲状腺激素。

Enzyme-free chemiluminescence immunoassay for the determination of thyroid stimulating hormone.

机构信息

Luminescent MD, LLC, Hagerstown, MD 21742, United States; College of Engineering, Cornell University, Ithaca, NY 14853, United States.

Luminescent MD, LLC, Hagerstown, MD 21742, United States.

出版信息

Talanta. 2017 Aug 15;171:229-235. doi: 10.1016/j.talanta.2017.05.007. Epub 2017 May 4.

Abstract

Based on the chemiluminescent resonance energy transfer (CRET) from high-energy intermediate formed in 1,1'-oxalyldiimidazole chemiluminescence (ODI-CL) reaction to fluorescent microsphere-conjugated detection antibody, a cost-effective and highly sensitive enzyme-free sandwich immunoassay with chemiluminescence detection was developed for the early diagnosis of thyroid cancer. Thyroid stimulating hormone (TSH) primary antibody immobilized on paramagnetic bead captures rapidly TSH in human serum within 15min. Also, TSH detection antibody-conjugated yellow-green microsphere, capable of emitting green light, rapidly interacted with TSH-bound TSH primary antibody immobilized on paramagnetic bead during the 30-min incubation at room temperature. The sandwich complexes immobilized on paramagnetic in PBS buffer (pH 7.4) emit bright green light with the addition of ODI-CL reagents (e.g., HO, ODI). The dynamic range of enzyme-free sandwich immunoassay was 0.037-18μIUml with 3-5% coefficient of variation. The limit of detection (LOD=background+3σ) of the enzyme free sandwich immunoassay was as low as 0.011μIUml. We expect that the enzyme free sandwich immunoassay with ODI-CL detection can be a new assay method for the early diagnosis of thyroid cancer with statistically acceptable accuracy, precision, and reproducibility.

摘要

基于 1,1'-草酰二脒基(ODI)化学发光反应中高能中间体向荧光微球偶联检测抗体的化学发光共振能量转移(CRET),建立了一种经济高效、高灵敏度的无酶三明治免疫分析法,用于甲状腺癌的早期诊断。甲状腺刺激素(TSH)抗体固定在顺磁性磁珠上,可在 15min 内快速捕获人血清中的 TSH。此外,TSH 检测抗体偶联的黄绿光微球能够在室温下孵育 30min 时与固定在顺磁性磁珠上的 TSH 结合的 TSH 一抗快速相互作用。在 PBS 缓冲液(pH 7.4)中固定在顺磁性上的三明治复合物在加入 ODI-CL 试剂(如 HO、ODI)后发出明亮的绿光。无酶三明治免疫分析法的动态范围为 0.037-18μIUml,变异系数为 3-5%。无酶三明治免疫分析法的检测限(LOD=背景+3σ)低至 0.011μIUml。我们预计,基于 ODI-CL 检测的无酶三明治免疫分析法具有统计学上可接受的准确性、精密度和重现性,可作为甲状腺癌早期诊断的新方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验