Katayama Minako, Zarbatany David, Cha Stephen S, Fatemi Mostafa, Belohlavek Marek
Department of Cardiovascular Diseases, Mayo Clinic, Scottsdale, Arizona, USA.
Independent Engineering Consultant, Laguna Niguel, California, USA.
Ultrasound Med Biol. 2017 Sep;43(9):1888-1896. doi: 10.1016/j.ultrasmedbio.2017.04.014. Epub 2017 Jun 5.
Navigation of intracardiac catheters by echocardiography is challenging because of the fundamental limitations of B-mode ultrasonography. We describe a catheter fitted with a piezoelectric crystal, which vibrates and produces an instantaneous marker in color flow Doppler scans. The navigation learning curve was explored first in six pigs. Accuracy and precision of targeting with the navigation marker "off" (i.e., B-mode imaging) and "on" were assessed in another six pigs. Paired comparisons confirmed significantly (p = 0.04) shorter mean distances achieved in each pig with the color Doppler marker. Pooled (mean ± standard deviation) distance of the catheter tip from the target crystal was 5.27 ± 1.62 mm by B-mode guidance and 3.66 ± 1.45 mm by color Doppler marker navigation. Dye injection targeted into the ischemic border zone was successful in 8 of 10 pigs. Intracardiac catheter navigation with color Doppler ultrasonography is more accurate compared with conventional guidance by B-mode imaging.
由于B型超声检查存在根本局限性,通过超声心动图引导心内导管操作具有挑战性。我们描述了一种装有压电晶体的导管,该晶体在彩色多普勒扫描中振动并产生瞬时标记。首先在六头猪身上探索了导航学习曲线。在另外六头猪身上评估了在导航标记“关闭”(即B型成像)和“开启”状态下靶向的准确性和精确性。配对比较证实,每头猪使用彩色多普勒标记时平均距离显著缩短(p = 0.04)。通过B型引导,导管尖端与目标晶体的合并(平均值±标准差)距离为5.27±1.62毫米,通过彩色多普勒标记导航为3.66±1.45毫米。在10头猪中,有8头成功将染料注入缺血边缘区。与传统的B型成像引导相比,彩色多普勒超声心动图引导心内导管导航更准确。