Clinical recognition of pacemaker battery depletion and automatic reprogramming.

作者信息

Sinha Sunil K, Chrispin Jonathan, Barth Andreas, Rickard John Jack, Spragg David D, Berger Ronald, Calkins Hugh, Tomaselli Gordon, Marine Joseph E

机构信息

Division of Cardiology, Johns Hopkins University School of Medicine, Baltimore, MD.

Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, OH.

出版信息

Pacing Clin Electrophysiol. 2017 Aug;40(8):969-974. doi: 10.1111/pace.13135. Epub 2017 Jul 12.

Abstract

All contemporary pacemakers undergo automatic reprogramming upon reaching elective replacement indication due to battery depletion. The majority of such reprogramming will result in changes to both pacing mode and pacing rate. The exact software reprogramming varies considerably among pacemaker manufacturers and may even vary among models of the same manufacturer. Accordingly, it is essential for healthcare providers managing pacemaker patients to have a detailed understanding of the automatic reprogramming seen at elective replacement indication as well as their potential physiological and clinical consequences.

摘要

所有当代起搏器在因电池耗尽达到择期更换指征时都会进行自动重新编程。大多数此类重新编程会导致起搏模式和起搏频率都发生变化。这种软件重新编程的确切情况在不同的起搏器制造商之间有很大差异,甚至在同一制造商的不同型号之间也可能有所不同。因此,对于管理起搏器患者的医疗保健提供者来说,详细了解在择期更换指征时出现的自动重新编程及其潜在的生理和临床后果至关重要。