Med Phys. 2017 Sep;44(9):4786-4794. doi: 10.1002/mp.12415. Epub 2017 Jul 20.
X-ray CT faces challenges in differentiating tumors from surrounding healthy tissues. A bioluminescence tomography (BLT) system which can directly reconstruct the internal luminescent tumors, was developed and integrated with CT system to accurately guide radiation dose delivery.
The BLT system, employing a lens-coupled CCD camera, was physically registered with an onboard cone beam CT system in an image-guided small animal arc radiation treatment system (iSMAART). The onboard CT provides animal anatomy and accurate surface contour used to construct the three-dimensional mesh for the BLT reconstruction. Bioluminescence projections were captured from multiple angles, once every 90° rotation. The BLT reconstruction was performed on an orthotopic prostate tumor model to evaluate its robustness and accuracy in locating and delineating bioluminescent tumors. The location and volume of the tumor identified from iodinated contrast CT was used to validate the BLT performance. Phantom experiment was also conducted to confirm the precision of BLT-guided radiation.
The BLT was able to accurately locate the bioluminescent tumors with < 0.5 mm error. The tumor volume in BLT was significantly correlated with that in the iodinated contrast CT (R = 0.81, P < 0.001). Phantom experiments further validated that BLT can be used to guide radiation with submillimeter accuracy.
Together with CT, BLT can provide precision radiation guidance and robust tumor volume assessment in small animal cancer research.
X 射线 CT 在区分肿瘤与周围健康组织方面面临挑战。本研究开发了一种可直接重建内部发光肿瘤的生物发光断层成像(BLT)系统,并将其与 CT 系统集成,以准确指导放射剂量输送。
BLT 系统采用透镜耦合 CCD 相机,与图像引导小动物弧形放射治疗系统(iSMAART)中的内置锥形束 CT 系统在物理上进行了配准。内置 CT 提供动物解剖结构和准确的表面轮廓,用于构建 BLT 重建的三维网格。生物发光投影从多个角度捕获,每次旋转 90°。在原位前列腺肿瘤模型上进行 BLT 重建,以评估其在定位和描绘生物发光肿瘤方面的稳健性和准确性。使用碘造影 CT 确定的肿瘤位置和体积来验证 BLT 性能。还进行了体模实验以确认 BLT 引导放射的精度。
BLT 能够准确定位生物发光肿瘤,误差小于 0.5 毫米。BLT 中的肿瘤体积与碘造影 CT 中的肿瘤体积显著相关(R = 0.81,P < 0.001)。体模实验进一步验证了 BLT 可用于以亚毫米精度引导放射。
BLT 与 CT 结合可在小动物癌症研究中提供精确的放射引导和稳健的肿瘤体积评估。