Rojas-González Antonio, Cecconi Alberto, Merino Jose L, Jimenez-Borreguero Luis J, Alfonso Fernando
Servicio de Cardiología, Hospital Universitario de La Princesa, Madrid, Spain.
Servicio de Cardiología, Hospital Universitario de La Paz, Madrid, Spain.
J Electrocardiol. 2017 Nov-Dec;50(6):937-938. doi: 10.1016/j.jelectrocard.2017.06.018. Epub 2017 Jun 8.
Dual-chamber pacing is feasible via the floating atrial sensor electrodes of a single-pass VDD lead but the atrial lead threshold is higher than the accepted clinical standard. Furthermore, due to the floating nature of the system, atrial sensing and pacing thresholds may vary during the follow up. For these reasons this strategy is seldom considered a common pacing solution in routine clinical practice. Alternatively, this phenomenon is likely to be observed as a result of incorrect generator configuration. As shown in our case, this inadequate setting can be suspected just by the analysis of the surface ECG and the post implantation chest X-ray.
通过单通道VDD导线的漂浮心房感知电极进行双腔起搏是可行的,但心房导线阈值高于公认的临床标准。此外,由于该系统的漂浮特性,心房感知和起搏阈值在随访期间可能会发生变化。由于这些原因,在常规临床实践中,这种策略很少被视为一种常见的起搏解决方案。或者,这种现象可能是由于发生器配置不正确而观察到的。如我们的病例所示,仅通过分析体表心电图和植入后胸部X线片就可以怀疑这种设置不当。