Suppr超能文献

一种模块化股骨柄设计中植入物骨折发生率高得惊人:两种植入物的比较。

Alarmingly High Rate of Implant Fractures in One Modular Femoral Stem Design: A Comparison of Two Implants.

作者信息

Shah Ritesh R, Goldstein Jeffrey M, Cipparrone Nancy E, Gordon Alexander C, Jimenez Matthew L, Goldstein Wayne M

机构信息

Department of Orthopedic Surgery, Illinois Bone & Joint Institute, Morton Grove, Illinois; Department of Orthopedic Surgery, Advocate Lutheran General Hospital, Park Ridge, Illinois; Department of Orthopedic Surgery, NorthShore University HealthSystem - Skokie Hospital, Skokie, Illinois.

Department of Orthopedic Surgery, Illinois Bone & Joint Institute, Morton Grove, Illinois.

出版信息

J Arthroplasty. 2017 Oct;32(10):3157-3162. doi: 10.1016/j.arth.2017.05.031. Epub 2017 May 22.

Abstract

BACKGROUND

Reports of implant fracture at the modular junction have been seen in modular neck designs, stem-sleeve modular femoral stems, and diaphyseal engaging bi-body modular stems. To date, however, there has never been a direct comparison between 2 different implant designs from the same modular family. The purpose of this study is to compare the rate of implant failure of 2 such stem-sleeve modular femoral stem designs, the S-ROM and Emperion, to further identify factors which increase the risk of this mode of failure.

METHODS

A retrospective, single surgeon, review of our institutional database was performed to compare the 2 groups of patients.

RESULTS

A total of 1168 total hip arthroplasty procedures were included in our analysis, 547 (47%) with Emperion and 621 (53%) with S-ROM. Eight (1.5%) fractures in 7 patients occurred in the Emperion group compared to 1 (0.2%) fracture in the S-ROM group (P = .015).

CONCLUSION

The precise cause of the stem fractures in our study remains unknown and is likely multifactorial. Given the unexpectedly high rate of catastrophic implant failures in the form of stem fracture at the stem-sleeve junction, we recommend more judicious use of modularity in primary total hip arthroplasty.

摘要

背景

在模块化颈设计、柄-套筒模块化股骨干以及骨干啮合双体模块化柄中,均可见到模块化连接处种植体骨折的报道。然而,迄今为止,同一模块化系列的两种不同种植体设计之间从未进行过直接比较。本研究的目的是比较两种此类柄-套筒模块化股骨干设计(S-ROM和Emperion)的种植体失败率,以进一步确定增加这种失败模式风险的因素。

方法

对我们机构数据库进行回顾性、单术者的分析,以比较两组患者。

结果

我们的分析共纳入1168例全髋关节置换手术,其中547例(47%)使用Emperion,621例(53%)使用S-ROM。Emperion组7例患者出现8处(1.5%)骨折,而S-ROM组有1处(0.2%)骨折(P = 0.015)。

结论

我们研究中柄骨折的确切原因尚不清楚,可能是多因素的。鉴于柄-套筒连接处柄骨折形式的灾难性种植体失败发生率出乎意料地高,我们建议在初次全髋关节置换术中更谨慎地使用模块化。