Rueckl Kilian, Boettner Friedrich, Bechler Ulrich, Baral Elexis C, Wright Timothy M, Sculco Peter K
Hospital for Special Surgery, New York, NY, USA.
Department for Orthopedic Surgery, University of Wuerzburg, Germany.
Arthroplast Today. 2018 May 10;4(3):295-299. doi: 10.1016/j.artd.2018.02.004. eCollection 2018 Sep.
Fracture of a well-ingrown femoral component is a rare and often challenging complication. Modular junctions and sleeve interfaces have been identified as one potential point of weakness with corrosion and fretting being contributing factors to ultimate femoral component fracture. Stem fractures at the sleeve interface were reported occasionally for the proximal ingrowth modular Emperion System (Smith and Nephew, Memphis, TN). However, this failure mechanism has been reported infrequently, often associated with corrosion at the modular junction, for the similarly designed S-ROM system (DePuy Orthopedics Inc., Warsaw, IN). We present the case of a 52-year-old patient, with a body weight of 84 kg (185 lbs) and a body mass index of 30.6 kg/m, who suffered a fatigue fracture of a 14 × 09 × 130 mm S-ROM stem 42 months after implantation. The present study presents the results of the surface analysis, discusses possible failure mechanisms, provides treatment guidelines, and a review of the literature revealing 15 cases of failure at the level of the stem-sleeve junction. In particular, modifiable risk factors for potential stem failure, including stem diameter, stem offset, and the resulting cantilever bending forces on the proximal sleeve-stem junction, are discussed in detail.
生长良好的股骨假体骨折是一种罕见且颇具挑战性的并发症。模块化连接和套筒界面已被确定为一个潜在的薄弱点,腐蚀和微动是导致最终股骨假体骨折的因素。对于近端生长型模块化Emperion系统(史赛克公司,田纳西州孟菲斯),偶尔会有关于套筒界面处柄部骨折的报道。然而,对于设计类似的S-ROM系统(迪普伊骨科公司,印第安纳州华沙),这种失效机制的报道较少,且常与模块化连接处的腐蚀有关。我们报告一例52岁患者,体重84千克(185磅),体重指数为30.6千克/平方米,在植入14×09×130毫米的S-ROM柄42个月后发生疲劳骨折。本研究展示了表面分析结果,讨论了可能的失效机制,提供了治疗指南,并对文献进行了综述,发现有15例在柄-套筒连接处出现失效的病例。特别是,详细讨论了潜在柄部失效的可改变风险因素,包括柄直径、柄偏移以及近端套筒-柄连接处产生的悬臂弯曲力。