Magn Reson Med Sci. 2018 Apr 10;17(2):161-167. doi: 10.2463/mrms.mp.2017-0037. Epub 2017 Jun 26.
We investigated the feasibility of single breath hold unenhanced coronary MRA using multi-shot gradient echo planar imaging (MSG-EPI) on a 3T-scanner.
Fourteen volunteers underwent single breath hold coronary MRA with a MSG-EPI and free-breathing turbo field echo (TFE) coronary MRA at 3T. The acquisition time, signal to noise ratio (SNR), and the contrast of the sequences were compared with the paired t-test. Readers evaluated the image contrast, noise, sharpness, artifacts, and the overall image quality.
The acquisition time was 88.1% shorter for MSG-EPI than TFE (24.7 ± 2.5 vs 206.4 ± 23.1 sec, P < 0.01). The SNR was significantly higher on MSG-EPI than TFE scans (P < 0.01). There was no significant difference in the contrast on MSG-EPI and TFE scans (1.8 ± 0.3 vs 1.9 ± 0.3, P = 0.24). There was no significant difference in image contrast, image sharpness, and overall image quality between two scan techniques. The score of image noise and artifact were significantly higher on MSG-EPI than TFE scans (P < 0.05).
The single breath hold MSG-EPI sequence is a promising technique for shortening the scan time and for preserving the image quality of unenhanced whole heart coronary MRA on a 3T scanner.
我们研究了在 3T 扫描仪上使用多回波梯度回波平面成像(MSG-EPI)单次屏气无增强冠状动脉 MRA 的可行性。
14 名志愿者在 3T 上进行了单次屏气冠状动脉 MRA 和自由呼吸涡轮场回波(TFE)冠状动脉 MRA。使用配对 t 检验比较了采集时间、信噪比(SNR)和序列对比度。读者评估了图像对比度、噪声、锐度、伪影和整体图像质量。
MSG-EPI 的采集时间比 TFE 缩短了 88.1%(24.7 ± 2.5 秒与 206.4 ± 23.1 秒,P < 0.01)。MSG-EPI 的 SNR 明显高于 TFE 扫描(P < 0.01)。MSG-EPI 和 TFE 扫描的对比度无显著差异(1.8 ± 0.3 与 1.9 ± 0.3,P = 0.24)。两种扫描技术的图像对比度、图像锐度和整体图像质量无显著差异。图像噪声和伪影评分在 MSG-EPI 上明显高于 TFE 扫描(P < 0.05)。
单次屏气 MSG-EPI 序列是一种很有前途的技术,可缩短扫描时间,并保持 3T 扫描仪上无增强全心冠状动脉 MRA 的图像质量。