The purpose of this study was to investigate the feasibility of non-contrast renal MRA using multi-shot gradient echo planar imaging (MSG-EPI) with a 3-T MRI system.
Seventeen healthy volunteers underwent non-contrast renal MRA using MSG-EPI and balanced steady-state free precession (b-SSFP) sequences on a 3-T MRI system. Two radiologists independently recorded the images' contrast, noise, sharpness, artifacts, and overall quality on 4-point scales. The signal-to-noise ratio (SNR) for the renal artery, the contrast ratio (CR) between the renal artery and erector spinae, and acquisition time were compared between the two sequences.
The SNR and CR were significantly higher with MSG-EPI than with the b-SSFP sequence (17.80 ± 3.67 vs. 10.84 ± 2.86 and 0.77 ± 0.05 and 0.66 ± 0.09, respectively; p < 0.05), and the acquisition time was significantly lower (164.5 ± 34.0 vs. 261.5 ± 39.3 s, respectively; p < 0.05). There were significant differences in image contrast, noise, sharpness, artifacts, and overall image quality between the two sequences (p < 0.01).
The MSG-EPI sequence is a promising technique that can shorten the scan time and improve the image quality of non-contrast renal MRA with a 3-T MRI system.
• The multi-shot gradient echo planar imaging with an inversion pulse is a brand-new fast scan technique for an unenhanced renal MRA. • The image quality of multi-shot gradient echo planar imaging is better than that of b-SSFP for an unenhanced renal MRA.
本研究旨在探讨使用3-T磁共振成像(MRI)系统的多次激发梯度回波平面成像(MSG-EPI)进行非对比剂肾磁共振血管造影(MRA)的可行性。
17名健康志愿者在3-T MRI系统上使用MSG-EPI和平衡稳态自由进动(b-SSFP)序列进行非对比剂肾MRA。两名放射科医生独立地以4分制记录图像的对比度、噪声、清晰度、伪影和整体质量。比较了两种序列之间肾动脉的信噪比(SNR)、肾动脉与竖脊肌之间的对比率(CR)以及采集时间。
MSG-EPI序列的SNR和CR显著高于b-SSFP序列(分别为17.80±3.67与10.84±2.86以及0.77±0.05与0.66±0.09;p<0.05),且采集时间显著更短(分别为164.5±34.0与261.5±39.3秒;p<0.05)。两种序列在图像对比度、噪声、清晰度、伪影和整体图像质量方面存在显著差异(p<0.01)。
MSG-EPI序列是一种有前景的技术,可缩短扫描时间并提高3-T MRI系统非对比剂肾MRA的图像质量。
• 带有反转脉冲的多次激发梯度回波平面成像是一种用于未增强肾MRA的全新快速扫描技术。• 对于未增强肾MRA,多次激发梯度回波平面成像的图像质量优于b-SSFP。