使用磁共振引导的高强度聚焦超声消融骶髂关节:猪模型的初步实验

Ablation of the sacroiliac joint using MR-guided high intensity focused ultrasound: a preliminary experiment in a swine model.

作者信息

Kaye Elena A, Maybody Majid, Monette Sebastien, Solomon Stephen B, Gulati Amitabh

机构信息

Department of Medical Physics, Memorial Sloan Kettering Cancer Center, 1275 York Ave, New York, NY 10065 USA.

Department of Radiology, Memorial Sloan Kettering Cancer Center, 1275 York Ave, New York, NY 10065 USA.

出版信息

J Ther Ultrasound. 2017 Jun 26;5:17. doi: 10.1186/s40349-017-0095-x. eCollection 2017.

Abstract

BACKGROUND

Dysfunction of the Sacroiliac Joint (SIJ) is one of the key sources of low back pain. For prolonged pain relief, some patients undergo fluoroscopic guided radio-frequency (RF) ablation of SIJ, during which a number of RF probes are inserted to create thermal lesions that disrupt the posterior sacral nerve supply. This procedure is minimally invasive, laborious, time-consuming and costly. To study if High Intensity Focused Ultrasound (HIFU) ablation is a feasible alternative approach to SIJ pain treatment, we performed experiments using HIFU to ablate SIJ in the swine model.

METHODS

Three female Yorkshire swine (36, 35.2 and 34 kg) underwent bilateral Magnetic Resonance guided HIFU (MRgHIFU) ablation of the SIJs. Treatment assessment was performed using contrast-enhanced imaging, histopathology and evaluation of pain and changes in ambulation and gait.

RESULTS

Contiguous lesions along the right and left SIJs were achieved in all animals. In one out of three animals, excessive heating of the muscle and skin tissue in the near-field resulted in unwanted muscle necrosis. No changes in animal behavior, ambulation or gait were detected.

CONCLUSIONS

The initial experiments with MRgHIFU ablation of SIJs in sub-acute swine model show promise for this ablation modality as a non invasive and more precise alternative to the currently used fluoroscopically - guided RF ablations and injections.

摘要

背景

骶髂关节功能障碍是下腰痛的主要原因之一。为了实现长期疼痛缓解,一些患者接受了荧光镜引导下的骶髂关节射频消融术,在此过程中会插入多个射频探头以形成热损伤,从而破坏骶后神经供应。该手术具有微创性,但操作繁琐、耗时且成本高昂。为了研究高强度聚焦超声(HIFU)消融术是否是治疗骶髂关节疼痛的可行替代方法,我们使用HIFU对猪模型的骶髂关节进行了消融实验。

方法

对三只雌性约克夏猪(体重分别为36、35.2和34千克)的双侧骶髂关节进行了磁共振引导下的HIFU(MRgHIFU)消融。使用对比增强成像、组织病理学以及对疼痛、行走和步态变化的评估来进行治疗评估。

结果

所有动物的左右骶髂关节均形成了连续的损伤。三只动物中有一只,近场肌肉和皮肤组织过度受热导致了不必要的肌肉坏死。未检测到动物行为、行走或步态的变化。

结论

在亚急性猪模型中使用MRgHIFU消融骶髂关节的初步实验表明,这种消融方式有望成为一种非侵入性且比目前使用的荧光镜引导下射频消融和注射更精确的替代方法。