Angle Orthod. 1979 Apr;49(2):11O-20.
A panoramic head positioning device, the S.C., was tested for magnification production of radiographs and for repositioning. Ten skeletal Class I dentulous skulls with uniform metal markers were examined. The purpose was to see if a skeletal analysis could be obtained from panoramic radiographs. Basal skeletal magnification varied anatomically both vertically and horizontally, but the magnification percentages and their ranges, for the most part, were calculable. The few exceptions were believed due to lack of proper alignment of the S.C. and the General Electric Panelipse. Proper alignment could also correct the right to left asymmetrical magnification differences. Means and standard errors from 290 skeletal measurements were calculated and could serve as standards for skeletal analysis using panoramic cephalometrics.
一种全景头颅定位装置,即S.C.,针对X线片的放大成像及重新定位进行了测试。对10个带有均匀金属标记物的恒牙列Ⅰ类骨骼头颅进行了检查。目的是观察能否从全景X线片中获得骨骼分析结果。基底骨骼放大率在垂直和水平方向上均存在解剖学差异,但在大多数情况下,放大率百分比及其范围是可计算的。少数例外情况被认为是由于S.C.与通用电气Panelipse未正确对齐所致。正确对齐还可纠正左右不对称的放大差异。计算了290次骨骼测量的均值和标准误差,这些结果可作为使用全景头影测量法进行骨骼分析的标准。