Hasegawa Yuya, Suzuki Yasutada, Kawakubo Susumu
Yamanashi Institute of Public Health and Environment.
Graduate School, Faculty of Engineering, University of Yamanashi.
Anal Sci. 2017;33(7):859-862. doi: 10.2116/analsci.33.859.
An on-site determination method for trace arsenic has been developed by collecting it as molybdenum blue (MB) in the presence of tetradecyldimethylbenzylammonium chloride on a mixed cellulose ester membrane filter and by measuring reflection absorbance (RA) of MB on the filter using a laboratory-made palm-top size reflection-absorbance colorimeter with a red light-emitting diode. The value of RA was proportional to the amount of arsenic up to 0.5 μg with a detection limit of 0.01 μg. The proposed method was successfully applied to soil extract and hot-spring water samples.