Okazaki Takuya, Wang Wenjing, Kuramitz Hideki, Hata Noriko, Taguchi Shigeru
Department of Environmental Biology and Chemistry, Graduate School of Science and Engineering for Research, University of Toyama, Gofuku 3190, Toyama, Japan.
Anal Sci. 2013;29(1):67-72. doi: 10.2116/analsci.29.67.
An improved molybdenum blue spectrophotometry using a soluble membrane filter and CaCO(3)-column was proposed for determining arsenic in drinking water supplied from ground water in the presence of phosphate. A 100 mL sample solution containing 0.5 - 10 μg arsenic was passed through a CaCO(3)-column to remove phosphate, arsenate (As(V)). Arsenite (As(III)) which was not retained on the column was oxidized to As(V). As(V) was converted into a heteropolymolybdenum blue anion. The blue anion was collected on a membrane filter as an ion-associate with n-dodecyltrimethylammonium ion by filtration. The filter was dissolved in 2 mL of 2-methoxyethanol. The absorbance of the solution was measured at 810 nm against a reagent blank. Total inorganic arsenic was determined by reducing As(V) to As(III) before the column treatment. The RSDs for 10 μg L(-1) of As(III) and As(V) were 2.9%. Phosphate 0.2 mg L(-1) (as P) and iron 0.1 mg L(-1) did not interfere with the determination of 10 μg L(-1) arsenic. The proposed method was successfully applied to ground waters.
提出了一种改进的钼蓝分光光度法,该方法使用可溶性膜过滤器和碳酸钙柱,用于在存在磷酸盐的情况下测定地下水供应的饮用水中的砷。将含有0.5 - 10μg砷的100 mL样品溶液通过碳酸钙柱以去除磷酸盐、砷酸盐(As(V))。未保留在柱上的亚砷酸盐(As(III))被氧化为As(V)。As(V)转化为杂多钼蓝阴离子。通过过滤将蓝色阴离子作为与正十二烷基三甲基铵离子的离子缔合物收集在膜过滤器上。将过滤器溶解在2 mL的2 - 甲氧基乙醇中。以试剂空白为参比,在810 nm处测量溶液的吸光度。通过在柱处理前将As(V)还原为As(III)来测定总无机砷。10μg L(-1)的As(III)和As(V)的相对标准偏差为2.9%。0.2 mg L(-1)(以P计)的磷酸盐和0.1 mg L(-1)的铁不干扰10μg L(-1)砷的测定。该方法成功应用于地下水的检测。