Langley Joe, Pancani Silvia, Kilner Karen, Reed Heath, Stanton Andy, Heron Nicola, Judge Simon, McCarthy Avril, Baxter Susan, Mazzà Claudia, McDermott Christopher J
a Lab4Living, Sheffield Hallam University , Sheffield , UK.
b National Institute for Health Research Devices for Dignity Healthcare Technology Cooperative, Sheffield Teaching Hospitals Foundation Trust , Sheffield , UK.
Ergonomics. 2018 Feb;61(2):329-338. doi: 10.1080/00140139.2017.1353137. Epub 2017 Jul 28.
Identify location and intensity of discomfort experienced by healthy participants wearing cervical orthoses.
Convenience sample of 34 healthy participants wore Stro II, Philadelphia, Headmaster, and AspenVista cervical orthoses for four-hour periods. Participants reported discomfort level (scale 0-6) and location.
Participants reported mean discomfort for all orthoses over the four-hour test between 'a little discomfort' and 'very uncomfortable' (mean discomfort score = 1.64, SD = 1.50). Seven participants prematurely stopped tests due to pain and six reported maximum discomfort scores. Significant linear increase in discomfort with duration of wear was found for all orthoses. Significantly less discomfort was reported with Stro II than Headmaster and Philadelphia. Age correlated with greater perceived discomfort. Orthoses differed in the location discomfort was experienced.
Existing cervical orthoses cause discomfort influenced by design and duration of wear with orthoses' design the more significant factor. This work informed the design of a new orthosis and future orthoses developments. Practitioner Summary: The purpose of this study was to gain greater knowledge about the discomfort caused by wearing of existing neck orthoses in order to inform the design and development of a new neck orthosis. This study gathers empirical data from a surrogate population and concludes that orthosis design is more influential than the duration of wear.
确定佩戴颈椎矫形器的健康参与者所经历不适的位置和强度。
34名健康参与者的便利样本佩戴Stro II、费城、校长和AspenVista颈椎矫形器各4小时。参与者报告不适程度(0 - 6级)和位置。
在4小时的测试中,参与者报告所有矫形器的平均不适程度介于“有点不适”和“非常不适”之间(平均不适评分为1.64,标准差为1.50)。7名参与者因疼痛提前停止测试,6名报告了最大不适评分。所有矫形器的不适程度均随佩戴时间呈显著线性增加。与校长和费城矫形器相比,佩戴Stro II时报告的不适明显更少。年龄与更高的感知不适相关。不同矫形器的不适体验位置有所不同。
现有的颈椎矫形器会导致不适,其受设计和佩戴时间影响,其中矫形器设计是更重要的因素。这项工作为新型矫形器的设计及未来矫形器的发展提供了参考。从业者总结:本研究的目的是更深入了解佩戴现有颈部矫形器所引起的不适,以便为新型颈部矫形器的设计和开发提供参考。本研究从替代人群中收集了实证数据,并得出矫形器设计比佩戴时间更具影响力的结论。