滑动髋螺钉的疲劳失效——临床与生物力学分析

Fatigue failure of the sliding hip screw - clinical and biomechanical analysis.

作者信息

Necas Libor, Hrubina Maros, Cibula Zoltan, Behounek Jiri, Krivanek Stanislav, Horak Zdenek

机构信息

a Jessenius Faculty of Medicine in Martin , Comenius University in Bratislava , Martin , Slovac Republic.

b Orthopaedic Clinic , University Hospital Martin , Martin , Slovac Republic.

出版信息

Comput Methods Biomech Biomed Engin. 2017 Sep;20(12):1364-1372. doi: 10.1080/10255842.2017.1363192. Epub 2017 Aug 9.

Abstract

The study was aimed at the clinical and biomechanical analyses of the sliding hip screw breakage with the use of finite element method. We have identified two patients with the sliding screw breakage. In the first patient, the biomechanical analysis revealed the reduced stress values σ not exceeding the yield limit or strength limit of the implant. The yield limit was exceeded in second one. Clinical and biomechanical analyses have demonstrated that adherence to technical requirements of the appropriate osteosynthesis implementation is the principal condition of timely healing since it prevents the material failure.

摘要

本研究旨在运用有限元方法对滑动髋螺钉断裂进行临床和生物力学分析。我们确定了两名滑动螺钉断裂的患者。在第一名患者中,生物力学分析显示应力值σ降低,未超过植入物的屈服极限或强度极限。在第二名患者中,屈服极限被超过。临床和生物力学分析表明,遵守适当骨接合术实施的技术要求是实现及时愈合的主要条件,因为它可防止材料失效。