Measurement of cerebral blood flow by intravenous xenon-133 technique and a mobile system. Reproducibility using the Obrist model compared to total curve analysis.

作者信息

Schroeder T, Holstein P, Lassen N A, Engell H C

出版信息

Neurol Res. 1986 Dec;8(4):237-42. doi: 10.1080/01616412.1986.11739761.

Abstract

The recent development of a mobile 10 detector unit, using i.v. Xenon-133 technique, has made it possible to perform repeated bedside measurements of cerebral blood flow (CBF). Test-retest studies were carried out in 38 atherosclerotic subjects, in order to evaluate the reproducibility of CBF level and side-to-side asymmetry. Data were analysed according to the Obrist model and the results compared with those obtained using a model correcting for the air passage artifact. Reproducibility was of the same order of magnitude as reported using stationary equipment. The side-to-side CBF asymmetry was considerably more reproducible than CBF level. Using a single detector instead of five regional values averaged as the hemispheric flow increased standard deviation of CBF level by 10-20%, while the variation in asymmetry was doubled. In optimal measuring conditions the two models revealed no significant differences, but in low flow situations the artifact model yielded significantly more stable results. The present apparatus, equipped with 3-5 detectors covering each hemisphere, offers the opportunity of performing serial CBF measurements in situations not otherwise feasible.