Pitarch-Motellón J, Sancho J V, Ibáñez M, Pozo O, Roig-Navarro A F
Research Institute for Pesticides and Water, Universitat Jaume I, E-12071, Castellón, Spain.
Bioanalysis Research Group, IMIM, Hospital del Mar, 08003, Barcelona, Spain.
J Chromatogr A. 2017 Sep 15;1515:172-178. doi: 10.1016/j.chroma.2017.08.006. Epub 2017 Aug 3.
An isotope dilution mass spectrometry (IDMS) method for the determination of selected endogenous anabolic androgenic steroids (EAAS) in urine by UHPLC-MS/MS has been developed using the isotope pattern deconvolution (IPD) mathematical tool. The method has been successfully validated for testosterone, epitestosterone, androsterone and etiocholanolone, employing their respective deuterated analogs using two certified reference materials (CRM). Accuracy was evaluated as recovery of the certified values and ranged from 75% to 108%. Precision was assessed in intraday (n=5) and interday (n=4) experiments, with RSDs below 5% and 10% respectively. The method was also found suitable for real urine samples, with limits of detection (LOD) and quantification (LOQ) below the normal urinary levels. The developed method meets the requirements established by the World Anti-Doping Agency for the selected steroids for Athlete Biological Passport (ABP) measurements, except in the case of androsterone, which is currently under study.
已开发出一种采用同位素模式反褶积(IPD)数学工具,通过超高效液相色谱-串联质谱法(UHPLC-MS/MS)测定尿液中选定内源性合成代谢雄激素类固醇(EAAS)的同位素稀释质谱(IDMS)方法。该方法已使用两种有证标准物质(CRM),通过各自的氘代类似物成功验证了睾酮、表睾酮、雄酮和本胆烷醇酮。准确性通过认证值的回收率进行评估,范围为75%至108%。在日内(n=5)和日间(n=4)实验中评估精密度,相对标准偏差(RSD)分别低于5%和10%。该方法也适用于实际尿液样本,检测限(LOD)和定量限(LOQ)低于正常尿液水平。除了目前正在研究的雄酮外,所开发的方法符合世界反兴奋剂机构为运动员生物护照(ABP)测量中选定类固醇所制定的要求。