INTRODUCTION

We report a case of a lost metal platelet from a radiofrequency ablation probe (VAPR VUE Radiofrequency System, Cool Pulse 90, DePuy, Synthes, Switzerland) in the shoulder joint during elective arthroscopic cuff repair. To the best of our knowledge, this kind of an incident during elective arthroscopy has not been described in the literature so far. In addition, we present an algorithm on how to deal with such an incident.

CASE REPORT

A 69-year-old woman underwent an arthroscopic subacromial decompression and rotator cuff repair for a torn supraspinatus tendon. While performing the subacromial decompression and after swapping the portals from lateral to posterior, the metal platelet of the electrocautery device got detached from the instrument and lost in the operation field. Several attempts to visualize the lost platelet with the camera failed. Finally, intraoperative fluoroscopic imaging was used to detect the platelet. To confirm the definitive whereabouts of the platelet, two spinal needles were positioned perpendicular to another under x-ray control, both pointing at the missing platelet. After determining the exact location, the platelet could finally be visualized with the camera and removed. Due to this incident, the operation time was extended extensively, and the patient as well as the theatre team was exposed to an unnecessary amount of radiation.

CONCLUSION

This report indicates that an extraordinary incident such as the detachment of a component of the arthroscopic equipment during surgery is possible and should be kept in mind by the surgeon. Therefore, we believe that it is essential to perform a test of integrity at least at the end of every operation. In addition, we are presenting an algorithm on how to deal with the situation of a lost foreign body during arthroscopy, which can be applied to any joint.

摘要

引言

我们报告了一例在择期关节镜下肩袖修复术中,射频消融探头(VAPR VUE射频系统,Cool Pulse 90,德普伊,辛迪斯,瑞士)的金属小片在肩关节内丢失的病例。据我们所知,迄今为止,文献中尚未描述过择期关节镜检查期间发生的此类事件。此外,我们还提出了一种处理此类事件的算法。

病例报告

一名69岁女性因冈上肌腱撕裂接受关节镜下肩峰下减压和肩袖修复术。在进行肩峰下减压并将操作通道从外侧换到后侧后,电灼设备的金属小片从器械上脱落并在手术区域内丢失。多次尝试用摄像头找到丢失的小片均未成功。最后,术中使用荧光透视成像来检测小片。为了确定小片的确切位置,在X线控制下将两根脊椎穿刺针相互垂直放置,两根针都指向丢失的小片。确定确切位置后,最终用摄像头看到了小片并将其取出。由于这一事件,手术时间大幅延长,患者和手术团队都受到了不必要的辐射。

结论

本报告表明,手术过程中关节镜设备部件脱落等特殊事件是可能发生的,外科医生应牢记这一点。因此,我们认为至少在每次手术结束时进行完整性测试至关重要。此外,我们还提出了一种在关节镜检查期间处理异物丢失情况的算法,该算法可应用于任何关节。