U.S. Food and Drug Administration, Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, Division of Applied Mechanics, Silver Spring, MD 20993, USA.
Confluent Medical Technologies, Fremont, CA 94539, USA.
Acta Biomater. 2017 Oct 15;62:385-396. doi: 10.1016/j.actbio.2017.08.029. Epub 2017 Aug 24.
A major limitation with current assessments of corrosion in metallic medical devices is the lack of correlation between in-vitro and in-vivo corrosion performance. Therefore, the objective of this study was to elucidate the relationship between pitting corrosion measured by breakdown potentials (E) in ASTM F2129 testing and corrosion resistance in-vivo. Four groups of Nitinol stents were manufactured using different processing methods to create unique surface properties. The stents were implanted into iliac arteries of minipigs for six months and explanted for corrosion analysis. Scanning electron microscopy and energy dispersive X-ray spectrometry analyses indicated that stents with a thick complex thermal oxide (420nm) and high corrosion resistance in-vitro (E=975±94mV) were free from detectable corrosion in-vivo and exhibited no changes in Ni/Ti ratio when compared to non-implanted controls. This result was also found in mechanically polished stents with a thin native oxide (4nm; E=767±226mV). In contrast, stents with a moderately thick thermal oxide (130nm) and low corrosion resistance in-vitro (E=111±63mV) possessed corrosion with associated surface microcracks in-vivo. In addition, Ni/Ti ratios in corroded regions were significantly lower compared to non-corroded adjacent areas on explanted stents. When stents were minimally processed (i.e. retained native tube oxide from the drawing process), a thick thermal oxide was present (399nm) with low in-vitro corrosion resistance (E=68±29mV) resulting in extensive in-vivo pitting. These findings demonstrate that functional corrosion testing combined with a detailed understanding of the surface characteristics of a Nitinol medical device can provide insight into in-vivo corrosion resistance.
Nitinol is a commonly used material in the medical device industry. However, correlations between surface processing of nitinol and in-vivo corrosion has yet to be established. Elucidating the link between in-vivo corrosion and pre-clinical characterization can aid in improved prediction of clinical safety and performance of nitinol devices. We addressed this knowledge gap by fabricating nitinol stents to possess distinct surface properties and evaluating their corrosion susceptibility both in-vitro and after six months of in-vivo exposure. Relationships between stent processing, surface characterization, corrosion bench testing, and outcomes from explanted devices are discussed. These findings highlight the importance of surface characterization in nitinol devices and provide in-vitro pitting corrosion levels that can induce in-vivo corrosion in nitinol stents.
目前对金属医疗设备腐蚀的评估存在一个主要局限性,即体外和体内腐蚀性能之间缺乏相关性。因此,本研究的目的是阐明通过 ASTM F2129 测试测量的点蚀腐蚀电位(E)与体内耐腐蚀性之间的关系。使用不同的加工方法制造了四组 Nitinol 支架,以形成独特的表面特性。将支架植入小型猪的髂动脉中 6 个月,然后取出进行腐蚀分析。扫描电子显微镜和能量色散 X 射线光谱分析表明,具有厚复合热氧化物(420nm)和高体外耐腐蚀性(E=975±94mV)的支架在体内无检测到腐蚀,并且与未植入的对照相比,Ni/Ti 比没有变化。与具有薄自然氧化物(4nm;E=767±226mV)的机械抛光支架相比,也得到了相同的结果。相比之下,具有中等厚度热氧化物(130nm)和低体外耐腐蚀性(E=111±63mV)的支架在体内存在与表面微裂纹相关的腐蚀。此外,在腐蚀区域的 Ni/Ti 比值明显低于植入支架上未腐蚀的相邻区域。当支架经过最小处理(即保留拉伸过程中原有的管氧化物)时,存在厚的热氧化物(399nm),体外耐腐蚀性低(E=68±29mV),导致广泛的体内点蚀。这些发现表明,功能腐蚀测试结合对 Nitinol 医疗器械表面特性的详细了解,可以深入了解体内耐腐蚀性。
Nitinol 是医疗器械行业中常用的材料。然而,Nitinol 的表面处理与体内腐蚀之间的相关性尚未建立。阐明体内腐蚀与临床前特征之间的联系,可以帮助更好地预测 Nitinol 器械的临床安全性和性能。我们通过制造具有独特表面特性的 Nitinol 支架来解决这一知识空白,并在体外和体内暴露六个月后评估它们的腐蚀敏感性。讨论了支架加工、表面特性、腐蚀台架测试以及植入设备的结果之间的关系。这些发现强调了 Nitinol 器械表面特性的重要性,并提供了可以在 Nitinol 支架中引起体内腐蚀的体外点蚀腐蚀水平。