Wright Alex J, Bohner Austin D, Bernhisel Ashlie A, Zaugg Brian, Barlow William R, Pettey Jeff H, Olson Randall J
Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
Am J Ophthalmol. 2017 Nov;183:107-110. doi: 10.1016/j.ajo.2017.08.012. Epub 2017 Aug 30.
To evaluate the effects of micropulse, long pulse, and continuous ultrasound on transverse ultrasound using Abbott Medical Optics' (AMO) WhiteStar Signature Pro with the Ellips FX handpiece.
In vitro laboratory study.
This study was conducted at the John A. Moran Eye Center Laboratory, University of Utah, Salt Lake City, Utah, USA. Porcine lenses were hardened in formalin for 2 hours and equilibrated in basic salt solution (BSS) over a 24-hour period. The lenses were then cubed in 2.0 × 2.0-mm pieces. These pieces were stored in BSS until the time of experimentation. The AMO WhiteStar Signature Pro machine (Abbott Medical Optics) with the Ellips FX handpiece and a 0.9-mm bent Dewey tip with a 30-degree bevel (Microsurgical Technology Inc) were used for phacoemulsification. Three runs of 20 lenses each were performed, measuring efficiency and chatter. Transverse ultrasound varied in the 3 runs and included continuous, 6 ms on/off micropulse, and 50 ms on/off long pulse.
Micropulse was more efficient than long pulse by 43% (P = .00003) and continuous by 42% (P = .000387). There were also less chatter events with micropulse than with long-pulse and continuous ultrasound. However, this difference did not reach significance.
The 6 ms on and 6 ms off micropulse transverse 3-dimensional ultrasound is more efficient and produces fewer chatter events than both long-pulse and continuous ultrasound.
使用雅培医疗光学公司(AMO)的配备Ellips FX手持探头的WhiteStar Signature Pro设备,评估微脉冲、长脉冲和连续超声对横向超声的影响。
体外实验室研究。
本研究在美国犹他州盐湖城犹他大学的约翰·A·莫兰眼科中心实验室进行。猪晶状体在福尔马林中硬化2小时,然后在基础盐溶液(BSS)中平衡24小时。接着将晶状体切成2.0×2.0毫米的小块。这些小块储存在BSS中直至实验时使用。使用配备Ellips FX手持探头的AMO WhiteStar Signature Pro机器(雅培医疗光学公司)以及带有30度斜面的0.9毫米弯曲杜威探头(显微外科技术公司)进行超声乳化。每次对20个晶状体进行三轮操作,测量效率和震颤情况。三轮操作中的横向超声有所不同,包括连续超声、6毫秒开/关的微脉冲以及50毫秒开/关的长脉冲。
微脉冲比长脉冲效率高43%(P = 0.00003),比连续超声效率高42%(P = 0.000387)。微脉冲产生的震颤事件也比长脉冲和连续超声少。然而,这种差异未达到显著水平。
6毫秒开、6毫秒关的微脉冲横向三维超声比长脉冲和连续超声更高效,且产生的震颤事件更少。