Reliability and Precision of the Nana Protocol to Assess Body Composition Using Dual Energy X-Ray Absorptiometry.

作者信息

Shiel Flinn, Persson Carl, Simas Vini, Furness James, Climstein Mike, Pope Rod, Schram Ben

机构信息

1 Bond University.

2 The University of Sydney.

出版信息

Int J Sport Nutr Exerc Metab. 2018 Jan 1;28(1):19-25. doi: 10.1123/ijsnem.2017-0174. Epub 2018 Jan 24.

Abstract

The Nana positioning protocol is widely used to position participants to minimize technical error when undertaking body composition scanning and analysis with a Dual energy X-Ray absorptiometry (DXA) machine. Once biological and technical errors are accounted for, the only variation in test-retest results is from statistical fluctuation or machine error. Therefore, the aim of this study is to assess the test-retest reliability of the Nana positioning protocol and establish the smallest real difference percentage (SRD%). A gender-balanced group of 30 participants (15 males, 15 females) underwent two scans in succession using the Nana positioning protocol, with repositioning between scans. Percentage change in mean with typical error, Intraclass Correlation Coefficients (ICC), and standard error measurement percentage (SEM%) were used to identify the test-retest reliability and error rate of these protocols. Additionally, SRD% was calculated to assess the point at which clinically important changes occurred in a participant. The reliabilities of the whole body and regional scans were excellent. Percentage change in mean ranged between 0.00% and 0.23%. High reproducibility of the Nana positioning protocol was evident through an ICC ranging between 0.966-1.000. Additionally, typical error, SEM%, and SRD% were all low. Interestingly, fat mass was associated with the largest fluctuations observed to be associated with any of the parameters assessed. When all sources of biological and technical errors have been accounted for, the Nana positioning protocol has excellent test-retest reliability and produces low SEM% and SRD%.

摘要

娜娜定位方案被广泛用于对参与者进行定位,以在使用双能X线吸收法(DXA)机器进行身体成分扫描和分析时将技术误差降至最低。一旦考虑到生物学和技术误差,重测结果中唯一的差异就来自统计波动或机器误差。因此,本研究的目的是评估娜娜定位方案的重测可靠性,并确定最小真实差异百分比(SRD%)。一组性别均衡的30名参与者(15名男性,15名女性)按照娜娜定位方案连续进行了两次扫描,两次扫描之间重新定位。使用均值的百分比变化及典型误差、组内相关系数(ICC)和标准误差测量百分比(SEM%)来确定这些方案的重测可靠性和误差率。此外,计算SRD%以评估参与者发生临床重要变化的点。全身和局部扫描的可靠性极佳。均值的百分比变化在0.00%至0.23%之间。通过0.966 - 至1.000的ICC可以明显看出娜娜定位方案具有高再现性。此外,典型误差、SEM%和SRD%均较低。有趣的是,观察到脂肪量与所评估的任何参数中最大的波动相关。当考虑到所有生物学和技术误差来源时,娜娜定位方案具有出色的重测可靠性,并且产生较低的SEM%和SRD%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验