Physiol Meas. 2017 Oct 31;38(11):L17-L27. doi: 10.1088/1361-6579/aa8c52.
Conditional trans-rectal stimulation of the pudendal nerve could provide a viable solution to treat hyperreflexive bladder in spinal cord injury. A set threshold of the amplitude estimate of the external anal sphincter surface electromyography (sEMG) may be used as the trigger signal. The efficacy of such a device should be tested in a large scale clinical trial. As such, a probe should remain in situ for several hours while patients attend to their daily routine; the recording electrodes should be designed to be large enough to maintain good contact while observing design constraints. The objective of this study was to arrive at a design for intra-anal sEMG recording electrodes for the subsequent clinical trials while deriving the possible recording and processing parameters.
Having in mind existing solutions and based on theoretical and anatomical considerations, a set of four multi-electrode probes were designed and developed. These were tested in a healthy subject and the measured sEMG traces were recorded and appropriately processed.
It was shown that while comparatively large electrodes record sEMG traces that are not sufficiently correlated with the external anal sphincter contractions, smaller electrodes may not maintain a stable electrode tissue contact. It was shown that 3 mm wide and 1 cm long electrodes with 5 mm inter-electrode spacing, in agreement with Nyquist sampling, placed 1 cm from the orifice may intra-anally record a sEMG trace sufficiently correlated with external anal sphincter activity.
The outcome of this study can be used in any biofeedback, treatment or diagnostic application where the activity of the external anal sphincter sEMG should be detected for an extended period of time.
条件性经直肠阴部神经刺激可能为治疗脊髓损伤后的反射性膀胱提供可行的解决方案。外部肛门括约肌表面肌电图(sEMG)幅度估计的设定阈值可作为触发信号。此类设备的功效应在大规模临床试验中进行测试。因此,当患者进行日常活动时,探针应在原位停留数小时;记录电极的设计应足够大,以在观察设计限制的同时保持良好的接触。本研究的目的是为后续临床试验设计用于经肛门内 sEMG 记录的电极,同时得出可能的记录和处理参数。
考虑到现有的解决方案,并基于理论和解剖学考虑,设计并开发了一组四个多电极探头。在健康受试者中对这些探头进行了测试,并记录和适当处理了测量的 sEMG 迹线。
结果表明,虽然比较大的电极记录的 sEMG 迹线与外部肛门括约肌收缩没有足够的相关性,但较小的电极可能无法保持稳定的电极组织接触。结果表明,与奈奎斯特采样一致的 3mm 宽、1cm 长、电极间距为 5mm 的电极,放置在距肛门 1cm 处,可经肛门内记录与外部肛门括约肌活动具有足够相关性的 sEMG 迹线。
本研究的结果可用于任何生物反馈、治疗或诊断应用,其中需要长时间检测外部肛门括约肌 sEMG 的活动。