[Revision of Calibration Method of CT-Number to Stopping-Power-Ratio Conversion for Treatment Planning of Particle Therapy].

作者信息

Kanematsu Nobuyuki, Mori Shinichiro, Inaniwa Taku

机构信息

Medical Physics Section, National Institute of Radiological Sciences Hospital, QST.

Department of Accelerator and Medical Physics, National Institute of Radiological Sciences, QST.

出版信息

Igaku Butsuri. 2017;37(1):34-37. doi: 10.11323/jjmp.37.1_34.

Abstract

A calibration method for CT-number to stopping-power-ratio conversion was recently proposed as a revision of the Japanese de facto standard method that has been used at particle therapy centers in Japan for over a decade. The revised method deals with 11 representative tissues of specific elemental composition and density, based on a latest compilation of standard tissue data. We report here how the revision was actually implemented into clinical practice. We applied the revised method to 7 CT-scanning conditions currently in use for treatment planning. For each condition, we derived CT numbers and stopping-power ratios of the representative tissues to constitute a polyline conversion function. We analyzed the change of target water-equivalent depth by the revision for 38 beams in treatment plans for 13 randomly sampled patients. The revision caused a mean change of +0.3 mm with a standard deviation of 0.4 mm. The maximum change was +1.2 mm or +0.5% of the depth, which may be clinically insignificant. The transition to the revised method was straightforward and would slightly improve the accuracy of the beam range in particle therapy.