Suppr超能文献

所有肩锁关节注射都应在影像引导下进行吗?

Should all acromioclavicular joint injections be performed under image guidance?

作者信息

Javed S, Sadozai Z, Javed A, Din A, Schmitgen G

机构信息

Trauma and Orthopaedic Department, Royal Blackburn Hospital, Blackburn, UK.

出版信息

J Orthop Surg (Hong Kong). 2017 Sep-Dec;25(3):2309499017731633. doi: 10.1177/2309499017731633.

Abstract

INTRODUCTION

Steroid and local anaesthetic injection to the acromioclavicular joint (ACJ) is a very common diagnostic and therapeutic procedure, which is often performed in the outpatient department. However, it can be difficult to localize this joint because of its small size, presence of osteophytes and variable morphology in the population. We performed a study to determine whether the use of an image intensifier (X-ray guidance), in theatre, improves the accuracy of this injection.

METHODS

This was a prospective study carried out between March 2014 and March 2015. The injections were performed by two senior orthopaedic surgeons. First, we clinically palpated the ACJ and marked the area over this point as A. Then, with the use of a needle and an image intensifier in a single plane, we identified the actual location of the ACJ and marked this point as B. We measured the distance between A and B in millimetres (mm) and determined the accuracy of the injections. Further analysis taking into account the ACJ capsular attachments was also performed.

RESULTS

In total, 45 patients and 50 injections were included in the study; five patients had repeated injections at different times. We found that only 12 injections (24%) were palpated to be correct with no discrepancies between A and B (95% confidence interval: 14-37%). For the remaining 38 injections (76%), the use of an image intensifier had significantly improved the accuracy of ACJ location ( p < 0.05). Taking the capsular attachments of the ACJ into consideration reduced the number of inaccurate injections to 27 (54%).

CONCLUSION

We recommend the use of an image intensifier (or ultrasound guidance) to accurately determine the location of the ACJ for steroid and local anaesthetic injections. This prevents an injection into the wrong place, which can lead to wrong diagnosis and/or suboptimal treatment.

摘要

引言

向肩锁关节(ACJ)注射类固醇和局部麻醉剂是一种非常常见的诊断和治疗方法,常在门诊进行。然而,由于该关节体积小、存在骨赘以及人群中形态各异,可能难以对其进行定位。我们开展了一项研究,以确定在手术室使用影像增强器(X线引导)是否能提高该注射的准确性。

方法

这是一项于2014年3月至2015年3月进行的前瞻性研究。注射由两位资深骨科医生进行。首先,我们临床触诊ACJ并将该点上方区域标记为A。然后,在单一平面使用针和影像增强器,我们确定ACJ的实际位置并将该点标记为B。我们以毫米(mm)为单位测量A与B之间的距离,并确定注射的准确性。还进行了考虑ACJ囊附着情况的进一步分析。

结果

该研究共纳入45例患者和50次注射;5例患者在不同时间进行了重复注射。我们发现只有12次注射(24%)通过触诊被确定为正确,A与B之间无差异(95%置信区间:14 - 37%)。对于其余38次注射(76%),使用影像增强器显著提高了ACJ定位的准确性(p < 0.05)。考虑到ACJ的囊附着情况,不准确注射的数量减少至27次(54%)。

结论

我们建议使用影像增强器(或超声引导)来准确确定ACJ的位置,以便进行类固醇和局部麻醉剂注射。这可防止注射到错误部位,从而导致错误诊断和/或治疗效果不佳。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验