Accurate calibration of laparoscopic cameras is essential for enabling many surgical visualization and navigation technologies such as the ultrasound-augmented visualization system that we have developed for laparoscopic surgery. In addition to accuracy and robustness, there is a practical need for a fast and easy camera calibration method that can be performed on demand in the operating room (OR). Conventional camera calibration methods are not suitable for the OR use because they are lengthy and tedious. They require acquisition of multiple images of a target pattern in its entirety to produce satisfactory result. In this work, we evaluated the performance of a single-image camera calibration tool (; Percieve3D, Coimbra, Portugal) featuring automatic detection of corner points in the image, whether partial or complete, of a custom target pattern. Intrinsic camera parameters of a 5-mm and a 10-mm standard Stryker laparoscopes obtained using rdCalib and the well-accepted OpenCV camera calibration method were compared. Target registration error (TRE) as a measure of camera calibration accuracy for our optical tracking-based AR system was also compared between the two calibration methods. Based on our experiments, the single-image camera calibration yields consistent and accurate results (mean TRE = 1.18 ± 0.35 mm for the 5-mm scope and mean TRE = 1.13 ± 0.32 mm for the 10-mm scope), which are comparable to the results obtained using the OpenCV method with 30 images. The new single-image camera calibration method is promising to be applied to our augmented reality visualization system for laparoscopic surgery.

摘要

腹腔镜摄像头的精确校准对于实现许多手术可视化和导航技术至关重要,比如我们为腹腔镜手术开发的超声增强可视化系统。除了准确性和鲁棒性外,实际还需要一种快速简便的摄像头校准方法,以便能在手术室(OR)按需进行。传统的摄像头校准方法不适合在手术室使用,因为它们冗长且繁琐。它们需要完整采集目标图案的多幅图像才能产生满意的结果。在这项工作中,我们评估了一种单图像摄像头校准工具(;Percieve3D,葡萄牙科英布拉) 的性能,该工具能自动检测自定义目标图案图像中的角点,无论图像是部分还是完整的。比较了使用rdCalib和广泛认可的OpenCV摄像头校准方法获得的5毫米和10毫米标准史赛克腹腔镜的相机内参。还比较了两种校准方法之间作为基于光学跟踪的增强现实(AR)系统摄像头校准精度度量的目标配准误差(TRE)。基于我们的实验,单图像摄像头校准产生了一致且准确的结果(5毫米腹腔镜的平均TRE = 1.18 ± 0.35毫米,10毫米腹腔镜的平均TRE = 1.13 ± 0.32毫米),这与使用30幅图像的OpenCV方法获得的结果相当。新的单图像摄像头校准方法有望应用于我们的腹腔镜手术增强现实可视化系统。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索