LKEB, Division of Image Processing, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.
Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.
J Magn Reson Imaging. 2018 May;47(5):1397-1405. doi: 10.1002/jmri.25863. Epub 2017 Sep 27.
Myocardial tissue characterization by MR T and extracellular volume (ECV) mapping has demonstrated clinical value. The modified Look-Locker inversion recovery (MOLLI) sequence is a standard mapping technique, but its quality can be negatively affected by motion.
To develop a robust motion correction method for T and ECV mapping.
Retrospective analysis of clinical data.
Fifty patients who were referred to cardiac MR exam for T mapping.
FIELD STRENGTH/SEQUENCE: 3.0T cardiac MRI with precontrast and postcontrast MOLLI acquisition of the left ventricle (LV).
A groupwise registration method based on principle component analysis (PCA) was developed to register all MOLLI frames simultaneously. The resulting T and ECV maps were compared to those from the original and motion-corrected MOLLI with pairwise registration, in terms of standard deviation (SD) error.
Paired variables were compared using the Wilcoxon signed-rank test.
The groupwise registration method demonstrated improved registration performance compared to pairwise registration, with the T SD error reduced from 31 ± 20 msec to 26 ± 15 msec (P < 0.05), and ECV SD error reduced from 4.1 ± 3.6% to 2.8 ± 2.0% (P < 0.05). In LV segmental analysis, the performance was particularly improved in lateral segments, which are most affected by motion. The running time of groupwise registration was significantly shorter than that of the pairwise registration, 17.5 ± 3.0 seconds compared to 43.5 ± 2.2 seconds (P < 0.05).
We developed an automatic, robust motion correction method for myocardial T and ECV mapping based on a new groupwise registration scheme. The method led to lower mapping error compared to the conventional pairwise registration method in reduced execution time.
3 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2018;47:1397-1405.
心脏组织的 T1 和细胞外容积(ECV)磁共振成像特征已证明具有临床价值。改良 Look-Locker 反转恢复(MOLLI)序列是一种标准的测绘技术,但它的质量可能会受到运动的负面影响。
开发一种用于 T1 和 ECV 测绘的稳健运动校正方法。
临床数据的回顾性分析。
50 名因 T 映射而被转介到心脏磁共振检查的患者。
磁场强度/序列:3.0T 心脏 MRI 具有左心室(LV)的预对比和后对比 MOLLI 采集。
开发了一种基于主成分分析(PCA)的分组配准方法,用于同时配准所有 MOLLI 帧。然后将原始 MOLLI 与经运动校正的 MOLLI 之间的 T1 和 ECV 映射进行配对配准,并比较标准偏差(SD)误差。
采用配对变量的 Wilcoxon 符号秩检验。
分组配准方法与配对配准相比,显示出更好的配准性能,T1 的 SD 误差从 31±20 毫秒降低到 26±15 毫秒(P<0.05),ECV 的 SD 误差从 4.1±3.6%降低到 2.8±2.0%(P<0.05)。在 LV 节段分析中,运动影响最大的外侧节段的性能得到了特别改善。分组配准的运行时间明显短于配对配准,分别为 17.5±3.0 秒和 43.5±2.2 秒(P<0.05)。
我们开发了一种基于新的分组配准方案的自动、稳健的心肌 T1 和 ECV 测绘运动校正方法。与传统的配对配准方法相比,该方法在缩短执行时间的同时降低了映射误差。
3 技术功效:第 1 阶段 J. Magn. Reson. Imaging 2018;47:1397-1405.