Genc S, Dervisoglu E, Erdem S, Arslan O, Aktan M, Omer B
Istanbul Faculty of Medicine, Department of Biochemistry, Istanbul University, Istanbul, Turkey.
Istanbul Faculty of Medicine, Department of Hematology, Istanbul University, Istanbul, Turkey.
Int J Lab Hematol. 2017 Dec;39(6):633-640. doi: 10.1111/ijlh.12717. Epub 2017 Oct 4.
The purpose was to evaluate the analytical performances of Sysmex XN 3000 and UniCel DxH 800 comparing the obtained results with manual counting and between each other. Also flagging capabilities of abnormal cells were compared for both analyzers.
Two thousand one hundred and forty-two whole-blood samples were analyzed for evaluation. The samples flagged due to blast, atypical lymphocyte (AL), immature granulocyte (IG), or nucleated red blood cells (NRBC) were microscopically reviewed (n=102).
The within-run and between-day coefficient of variations (CV%) of XN 3000 for hemoglobin, RBC, MCV, WBC, and platelets were lower than 5% and for WBC differentials lower than 10% except monocytes which was 15.6% at low level. The precision results of DxH 800 were also lower than 5.0% except platelets (9.5%) and monocytes (45%) at low level. The comparison of analyzers revealed good agreement (R>.86), except monocytes and basophils. The flagging sensitivities of XN 3000 were higher for IGs, blasts, and ALs than those of DxH 800 and almost similar for NRBC.
The XN 3000 and DxH 800 are accurate, highly precise systems and can be used effectively in high-volume laboratories. The flagging sensitivity of XN 3000 was higher in detecting blasts, IGs, and ALs than that of DxH 800. The detection of abnormal cells with high sensitivity may improve laboratory workflow with a reduced slide review and accelerated turnaround time.
目的是评估Sysmex XN 3000和UniCel DxH 800的分析性能,将所得结果与手工计数结果进行比较,并比较两者之间的结果。同时还比较了两种分析仪对异常细胞的标记能力。
分析2142份全血样本进行评估。对因原始细胞、非典型淋巴细胞(AL)、未成熟粒细胞(IG)或有核红细胞(NRBC)而被标记的样本进行显微镜复查(n = 102)。
XN 3000血红蛋白、红细胞、平均红细胞体积、白细胞和血小板的批内和日间变异系数(CV%)低于5%,白细胞分类的变异系数低于10%,但单核细胞在低水平时为15.6%。DxH 800的精密度结果除血小板(9.5%)和单核细胞在低水平时(45%)外也低于5.0%。分析仪之间的比较显示出良好的一致性(R>.86),但单核细胞和嗜碱性粒细胞除外。XN 3000对IG、原始细胞和AL的标记敏感性高于DxH 800,对NRBC的标记敏感性几乎相似。
XN 3000和DxH 800是准确、高精度的系统,可有效地用于大容量实验室。XN 3000在检测原始细胞、IG和AL方面的标记敏感性高于DxH 800。高灵敏度检测异常细胞可减少玻片复查并加快周转时间从而改善实验室工作流程。