Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.

作者信息

Rastogi Shubham, Chandran Nair Sukesh, Murugan Pandiyan, Sukumar Asady Sukanya, Mammen Joy J, Mullai Saravanan

机构信息

HORIBA Medical, HORIBA ABX SAS, Parc Euromédecine - Rue du Caducée, France.

Department of Transfusion Medicine & Immunohematology, Christian Medical College, Vellore, Tamil Nadu, India.

出版信息

Pract Lab Med. 2022 Apr 18;30:e00275. doi: 10.1016/j.plabm.2022.e00275. eCollection 2022 May.

Abstract

This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over's K2 EDTA whole blood samples were analyzed for complete blood count. Of 400 samples, 200 were tested on Yumizen H2500; DxH 800 & DxH 900 while the other 200 were tested on Yumizen H2500 & XN-9000 (XN-10), respectively. The OEE was good and comparable for all the hematology analyzers except DxH 800 showing an average status. The sensitivity (%), specificity (%) and turnaround time (in minutes) for Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10) were 91.67, 61.11 & 103; 66.67, 54.55, & 149; 83.33, 27.27 & 136; 83.33, 28.57 & 122, respectively. Confusion matrix highlights the difficulty for DxH 800 and DxH 900 to discriminate left shift or blasts with large hyper-segmented neutrophils. The flags triggered by Yumizen H2500 were markedly changed to large hyper-segmented neutrophils. Lymphoblast caused more confusion for XN-9000 (XN-10), as it came out to be atypical lymphocytes, or hypersegmented neutrophils. Although comparable in OEE index to other analyzers, the Yumizen H2500 seems to be more reliable in detecting the abnormal cells as it has high sensitivity, specificity and less turnaround time. Thus, analysis adding specificity, sensitivity, and efficiency parameters to the OEE index provides more reliable information of the analyzers.

摘要

本研究评估并比较了高端血液分析仪Yumizen H2500、DxH 800、DxH 900和XN - 9000(XN - 10)的整体设备效能(OEE)、灵敏度、特异性和效率。总共对400份匿名剩余的K2 EDTA全血样本进行了全血细胞计数分析。在400份样本中,200份分别在Yumizen H2500、DxH 800和DxH 900上进行检测,另外200份分别在Yumizen H2500和XN - 9000(XN - 10)上进行检测。除DxH 800显示为平均水平外,所有血液分析仪的OEE良好且具有可比性。Yumizen H2500、DxH 800、DxH 900和XN - 9000(XN - 10)的灵敏度(%)、特异性(%)和周转时间(分钟)分别为91.67、61.11和103;66.67、54.55和149;83.33、27.27和136;83.33、28.57和122。混淆矩阵突出显示了DxH 800和DxH 900在区分左移或有大型分叶过多中性粒细胞的原始细胞方面存在困难。Yumizen H2500触发的标记明显变为大型分叶过多中性粒细胞。淋巴细胞白血病对XN - 9000(XN - 10)造成的混淆更多,因为它表现为非典型淋巴细胞或分叶过多中性粒细胞。尽管Yumizen H2500在OEE指数上与其他分析仪相当,但由于其具有高灵敏度、特异性且周转时间较短,似乎在检测异常细胞方面更可靠。因此,在OEE指数中加入特异性、灵敏度和效率参数进行分析,能提供关于分析仪更可靠的信息。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/5665/9117814/fc76f1a40129/gr1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验