Suppr超能文献

Fastrach 喉罩与 Spritztube 用于气道管理:一项基于模拟人的前瞻性随机研究。

Airway management with Fastrach laryngeal mask versus Spritztube: a prospective randomized manikin-based study.

机构信息

Department of Anesthesiology and Intensive Care, San Bortolo Hospital, Vicenza, Italy -

Department of Anesthesiology and Intensive Care, Sacro Cuore Catholic University, Rome, Italy -

出版信息

Minerva Anestesiol. 2018 Apr;84(4):455-462. doi: 10.23736/S0375-9393.17.11702-5. Epub 2017 Oct 4.

Abstract

BACKGROUND

A new promising device, the Spritztube (ST), was developed combining the ability to perform both supraglottic ventilation and orotracheal fibreoptic intubation using the same device, allowing an easy passage from supraglottic ventilation to tracheal ventilation avoiding apnea. The present study aims to compare the speed and the subjective ease of insertion of the novel tracheal tube (Spritztube®) compared to the intubating laryngeal mask airway Fastrach™ (FT-LMA) in a simulation environment.

METHODS

Each participant received verbal instruction and practical demonstration concerning "technique of insertion" for both devices on manikin and, in a randomized order, used both devices. Time of placement (T1), time of inflation (T2), the elapsed procedural time (T3), ease of insertion, time of exchange maneuver for intubation (T4), success rates and number of attempts were recorded for each EAD.

RESULTS

Forty-seven participants were enrolled. The ST was judged as easier insertion than FT-LMA (P<0.001) having also a significant higher success rate than FT-LMA (P<0.001). Number of attempts for insertion was similar. Number of attempts for exchange to ETI was significantly more for FT-LMA (P<0.001). FT-LMA was applied 11 s faster than the ST (median T3 Fastrach: 13 s, ST: 24 s, P<0.001) as the exchange maneuver for ETI (median T4 FT-LMA: 13 s, ST: 24 s, P<0.001).

CONCLUSIONS

In a manikin simulation setting, insertion and intubation with and ST was performed 11 s slower than with use of a FT-LMA.

摘要

背景

一种新的有前途的设备,Spritztube(ST),结合了既能进行声门上通气又能用同一设备进行经口纤维光导插管的能力,允许从声门上通气轻松过渡到气管通气,避免了窒息。本研究旨在比较新型气管导管(Spritztube®)与插管喉罩气道 Fastrach™(FT-LMA)在模拟环境中的插入速度和主观易插性。

方法

每位参与者均接受了关于两种设备的“插入技术”的口头指导和实践演示,然后以随机顺序使用两种设备。记录放置时间(T1)、充气时间(T2)、程序进行时间(T3)、插入的容易程度、插管交换操作时间(T4)、成功率和尝试次数。

结果

共纳入 47 名参与者。ST 被认为比 FT-LMA 更容易插入(P<0.001),并且其成功率也明显高于 FT-LMA(P<0.001)。插入的尝试次数相似。用于 ETI 插入的交换操作的尝试次数明显更多的是 FT-LMA(P<0.001)。FT-LMA 的应用速度比 ST 快 11 秒(FT-LMA 中位数 T3:13 秒,ST:24 秒,P<0.001),用于 ETI 的交换操作(FT-LMA 中位数 T4:13 秒,ST:24 秒,P<0.001)。

结论

在模拟人体模型设置中,与使用 FT-LMA 相比,使用 ST 进行插入和插管的速度慢 11 秒。