Ghisetti Valeria, Coignard Catherine, Allice Tiziano, Thoai Duong Ly
Laboratory Microbiology, Amedeo di Savoia Hospital, Turin, Italy.
Laboratory Eurofins Biomnis, Ivry sur Seine, France.
J Clin Virol. 2017 Nov;96:80-83. doi: 10.1016/j.jcv.2017.09.008. Epub 2017 Sep 21.
More and more countries test for HIV infection using combination assays that simultaneously detect p24 antigen and HIV antibodies.
To assess the performance of a new HIV combo assay: LIAISON XL murex HIV Ab/Ag HT.
The assays were examined with a total of 3090 samples that included 769 selected HIV antibody-negative samples, 1849 unselected HIV samples, 15 HIV-1 p24 Ag reference samples, 90 primary HIV-1 infection (PHI) samples, 167 HIV-1 antibody-positive samples (well characterized of groups M and O), 95 HIV-1 antibody-positive samples and 105 HIV-2 antibody-positive samples.
The specificity of the LIAISON XL murex HIV Ab/Ag HT was 99.7%. The analytical sensitivity of Ag p24 of the LIAISON XL murex HIV Ab/Ag HT was 0.58IU/mL and 9.93pg/mL when using WHO and French national standards, respectively. All screened HIV subtypes was identified by this assay. Also, 90 PHI specimens were detected by this screening assay.
The sensitivity and specificity of the LIAISON XL murex HIV Ab/Ag HT assay are high. Hence the assay offers automated high-throughput screening with ability to detect primary infection.
越来越多的国家使用同时检测p24抗原和HIV抗体的联合检测法来检测HIV感染。
评估一种新型HIV联合检测法——LIAISON XL murex HIV Ab/Ag HT的性能。
使用总共3090份样本对该检测法进行检测,这些样本包括769份经挑选的HIV抗体阴性样本、1849份未经挑选的HIV样本、15份HIV-1 p24抗原参考样本、90份原发性HIV-1感染(PHI)样本、167份HIV-1抗体阳性样本(M组和O组特征明确)、95份HIV-1抗体阳性样本以及105份HIV-2抗体阳性样本。
LIAISON XL murex HIV Ab/Ag HT的特异性为99.7%。当分别使用世界卫生组织和法国国家标准时,LIAISON XL murex HIV Ab/Ag HT的p24抗原分析灵敏度分别为0.58IU/mL和9.93pg/mL。该检测法能够识别所有筛查的HIV亚型。此外,该筛查检测法还检测出了90份PHI样本。
LIAISON XL murex HIV Ab/Ag HT检测法的灵敏度和特异性都很高。因此,该检测法可提供自动化高通量筛查,并具备检测原发性感染的能力。