COMPARISON BETWEEN PADC AND FNTD NEUTRON DETECTOR SYSTEMS IN BLIND TESTS.

作者信息

Yukihara E G, Akselrod M S, Fomenko V, Harrison J, Million M, Assenmacher F, Stabilini A, Meier K

机构信息

Department of Radiation Safety and Security, Paul Scherrer Institute, Villigen PSI, Switzerland.

Physics Department, Oklahoma State University, Stillwater, OK, USA.

出版信息

Radiat Prot Dosimetry. 2018 Aug 1;180(1-4):225-229. doi: 10.1093/rpd/ncx171.

Abstract

The objective of this study was to compare a neutron dosimetry system based on polyallyl diglycol carbonate (PADC) detectors with a new system based on Al2O3:C,Mg fluorescence nuclear track detectors (FNTD). The irradiations, performed as part of an intercomparison organized by the Physikalisch-Technische Bundesanstalt (PTB), Germany, were on a PMMA phantom with 252Cf or 241Am-Be source, usually with the phantom surface perpendicular to the radiation beam (0° angle), and with Hp(10) values between 0.3 and 7 mSv. One 252Cf irradiation was performed at 30° angle, and one with an additional 1 mSv gamma irradiation. The results showed an agreement between the two techniques with an average and maximum difference between PADCs and FNTDs of 1.5 and 22%, respectively, if one compares only cases of doses >1 mSv. For one of the irradiation conditions with dose of 0.9 mSv, use of the incorrect calibration factor for the FNTD (252Cf instead of 241Am-Be) led to reported values ~×2 larger than the given doses, due to low statistics in the determination of the ratio between 6Li-doped glass and polyethylene neutron converters. Although the FNTD track analysis algorithm may need further development, the results presented here demonstrate the feasibility of the FNTD technology and indicate areas requiring improvements.

摘要

相似文献

1
COMPARISON BETWEEN PADC AND FNTD NEUTRON DETECTOR SYSTEMS IN BLIND TESTS.
Radiat Prot Dosimetry. 2018 Aug 1;180(1-4):225-229. doi: 10.1093/rpd/ncx171.
2
HIGH DOSE FAST NEUTRON DOSIMETRY USING PADC PLASTIC NUCLEAR TRACK DETECTORS AND GREY LEVEL ANALYSIS.
Radiat Prot Dosimetry. 2018 Aug 1;180(1-4):220-224. doi: 10.1093/rpd/ncx186.
3
COMPARISON OF PADC NEUTRON DETECTORS FROM DIFFERENT SUPPLIERS.
Radiat Prot Dosimetry. 2018 Aug 1;180(1-4):196-200. doi: 10.1093/rpd/ncx193.
4
COMPARISON OF DIFFERENT PADC MATERIALS AND ETCHING CONDITIONS FOR FAST NEUTRON DOSIMETRY.
Radiat Prot Dosimetry. 2016 Sep;170(1-4):162-7. doi: 10.1093/rpd/ncv421. Epub 2015 Oct 5.
5
Automatic neutron dosimetry system based on fluorescent nuclear track detector technology.
Radiat Prot Dosimetry. 2014 Oct;161(1-4):86-91. doi: 10.1093/rpd/nct293. Epub 2013 Nov 26.
6
ENERGY RESPONSE OF FLUORESCENT NUCLEAR TRACK DETECTORS OF VARIOUS COLORATIONS TO MONOENERGETIC NEUTRONS.
Radiat Prot Dosimetry. 2018 Aug 1;180(1-4):215-219. doi: 10.1093/rpd/ncx232.
7
Simulation and calibration of an active neutron dosemeter.
Radiat Prot Dosimetry. 2014 Oct;161(1-4):126-9. doi: 10.1093/rpd/nct316. Epub 2013 Dec 23.
8
Novel Al2O3:C,Mg fluorescent nuclear track detectors for passive neutron dosimetry.
Radiat Prot Dosimetry. 2007;126(1-4):278-83. doi: 10.1093/rpd/ncm058. Epub 2007 May 23.
9
Brazilian gamma-neutron dosemeter: response to 241AmBe and 252Cf neutron sources.
Radiat Prot Dosimetry. 2011 Mar;144(1-4):215-7. doi: 10.1093/rpd/ncq550. Epub 2010 Dec 24.
10
Comparison of different PADC materials for neutron dosimetry.
Radiat Prot Dosimetry. 2014 Oct;161(1-4):104-7. doi: 10.1093/rpd/nct241. Epub 2013 Oct 3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验