通过离子色谱/低分辨率多接收器电感耦合等离子体质谱法直接测定δ钙同位素比值。

Direct determination of δ Ca isotope ratio by ion chromatography/low-resolution multicollector ICPMS.

作者信息

Karasinski Jakub, Bulska Ewa, Halicz Ludwik, Wojciechowski Marcin, Krata Agnieszka Anna

机构信息

Faculty of Chemistry, Biological and Chemical Research Centre, University of Warsaw, Żwirki i Wigury 101, 02-089, Warsaw, Poland.

Geological Survey of Israel, 30 Malkhei Israel St., Jerusalem, 95501, Israel.

出版信息

J Mass Spectrom. 2018 Jan;53(1):78-82. doi: 10.1002/jms.4039.

Abstract

High-precision on-line procedure for measurement of calcium isotopic ratio by coupling ion chromatography to multicollector inductively coupled plasma mass spectrometry was developed. Calcium separation from the sample matrix was achieved on an ion chromatography column-IonPac CS16-ID 3 mm connected with CERS 500 2 mm suppressor and followed by multicollector inductively coupled plasma mass spectrometry calcium isotopic ratio determination. Dry plasma mode was used with Aridus II desolvation system. To sustained samples with high level of total dissolved salts as well as account capacity of applied analytical column, the method has been optimized regarding calcium isotope ratio measurements with low-resolution mass spectrometry. Mass discrimination and instrument drift were corrected by sample-standard bracketing method using the Ca/ Ca isotope ratio of SRM 915a as a standard. Good accuracy and reasonable precision of calcium isotope ratio (generally 0.20‰ [2SD]) were achieved, which are comparable to off-line Ca separation and continuous measurement. The reproducibility of the proposed analytical procedure was verified by measuring the SRM 915a standard as a sample randomly over 3 months (n = 56). Applicability of the protocol was demonstrated for matrix-rich natural water samples, coral samples, and bone standard reference materials.

摘要

建立了一种将离子色谱与多接收电感耦合等离子体质谱联用测定钙同位素比值的高精度在线方法。在与CERS 500 2 mm抑制器相连的IonPac CS16-ID 3 mm离子色谱柱上实现了样品基质中钙的分离,随后进行多接收电感耦合等离子体质谱法测定钙同位素比值。采用Aridus II去溶剂化系统的干等离子体模式。为了处理总溶解盐含量高的样品以及考虑所用分析柱的容量,该方法针对低分辨率质谱法的钙同位素比值测量进行了优化。采用以SRM 915a的Ca/Ca同位素比值为标准的样品-标准品夹心法校正质量歧视和仪器漂移。获得了良好的准确度和合理的精密度(钙同位素比值一般为0.20‰ [2SD]),与离线钙分离和连续测量相当。通过在3个月内随机将SRM 915a标准品作为样品进行测量(n = 56),验证了所提出分析方法的重现性。证明了该方法对富含基质的天然水样、珊瑚样品和骨标准参考物质的适用性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验