Wright S M, Wright R M
Department of Radiology, Saint Francis Medical Center, Peoria, Ill.
Radiology. 1989 Jan;170(1 Pt 1):249-55. doi: 10.1148/radiology.170.1.2909105.
The authors describe a system of bilateral mutually coupled receiver coils and its initial clinical use in magnetic resonance imaging of the knees and temporomandibular joints. Switched mutually coupled coils allow bilateral simultaneous imaging of paired structures without a penalty in signal-to-noise ratio because the coils can be alternated with each section acquisition. With this coil system, there is no restriction on placement or design of coils to prevent interaction. When combined with three-dimensional imaging techniques, switched coils allow simultaneous bilateral acquisition of multiple, contiguous sections through paired anatomic structures in a reasonable amount of imaging time.
作者描述了一种双侧相互耦合的接收线圈系统及其在膝关节和颞下颌关节磁共振成像中的初步临床应用。开关式相互耦合线圈允许对成对结构进行双侧同时成像,而不会降低信噪比,因为每个层面采集时线圈可以交替使用。使用这种线圈系统,对线圈的放置或设计没有限制以防止相互干扰。当与三维成像技术结合使用时,开关式线圈能够在合理的成像时间内通过成对的解剖结构同时双侧采集多个连续层面。