Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery.

作者信息

Pälve H, Vuori A

机构信息

Department of Anesthesiology, University Central Hospital of Turku, Finland.

出版信息

Crit Care Med. 1989 Jan;17(1):66-9.

PMID:2909323
Abstract

Two commercially available pulse oximeters were found to read reliably early during periods of low cardiac index (CI) and low peripheral temperature (Tp). The lowest mean CI and mean Tp values at which a reading could be obtained were 2.4 L/min.m2 and 26.5 degrees C, respectively. The highest mean systemic vascular resistance index was 2930 dyne.sec/cm5.m2. The reliability of one monitor was equally good both in a normal and low perfusion mode (i.e., when the monitor displayed a low-quality signal message). The use of locally applied vasodilating creme widened the physiologic limits within which reliable measurements could be obtained. All oximeters and sensors overestimated slightly the saturation as compared to the hemoximeter readings, possibly because of a high mean carboxyhemoglobin concentration that was found to be present.