Fernandez Marco, Burns Kathy, Calhoun Beverly, George Saramma, Martin Beverly, Weaver Chris
Saint Thomas Health Services, Nashville, TN 37202, USA.
Am J Crit Care. 2007 Mar;16(2):146-52.
A new forehead noninvasive oxygen saturation sensor may improve signal quality in patients with low cardiac index.
To examine agreement between oxygen saturation values obtained by using digit-based and forehead pulse oximeters with arterial oxygen saturation in patients with low cardiac index.
A method-comparison study was used to examine the agreement between 2 different pulse oximeters and arterial oxygen saturation in patients with low cardiac index. Readings were obtained from a finger and a forehead sensor and by analysis of a blood sample. Bias, precision, and root mean square differences were calculated for the digit and forehead sensors. Differences in bias and precision between the 2 noninvasive devices were evaluated with a t test (level of significance P<.05).
Nineteen patients with low cardiac index (calculated as cardiac output in liters per minute divided by body surface area in square meters; mean 1.98, SD 0.34) were studied for a total of 54 sampling periods. Mean (SD) oxygen saturations were 97% (2.4) for blood samples, 96% (3.2) for the finger sensor, and 97% (2.8) for the forehead sensor. By Bland Altman analysis, bias +/- precision was -1.16 +/- 1.62% for the digit sensor and -0.36 +/- 1.74% for the forehead sensor; root mean square differences were 1.93% and 1.70%, respectively. Bias and precision differed significantly between the 2 devices; the forehead sensor differed less from the blood sample.
In patients with low cardiac index, the forehead sensor was better than the digit sensor for pulse oximetry.
一种新型前额无创血氧饱和度传感器可能会改善低心指数患者的信号质量。
研究低心指数患者使用指端脉搏血氧仪和前额脉搏血氧仪测得的血氧饱和度值与动脉血氧饱和度之间的一致性。
采用方法比较研究来检测低心指数患者中两种不同脉搏血氧仪与动脉血氧饱和度之间的一致性。从手指和前额传感器获取读数,并通过分析血样获得读数。计算指端和前额传感器的偏差、精密度和均方根差。用t检验评估两种无创设备在偏差和精密度上的差异(显著性水平P<0.05)。
对19例低心指数患者(计算方法为每分钟心输出量升数除以体表面积平方米数;均值1.98,标准差0.34)进行了总共54个采样期的研究。血样的平均(标准差)血氧饱和度为97%(2.4),手指传感器为96%(3.2),前额传感器为97%(2.8)。通过Bland Altman分析,指端传感器的偏差±精密度为-1.16±1.62%,前额传感器为-0.36±1.74%;均方根差分别为1.93%和1.70%。两种设备在偏差和精密度上有显著差异;前额传感器与血样的差异较小。
在低心指数患者中,前额传感器在脉搏血氧测定方面优于指端传感器。