Department of Gastroenterological and Pediatric Surgery, Oita University Faculty of Medicine, 1-1 Hasama-machi, Yufu, Oita, 879-5593, Japan.
TOGO MEDIKIT Co., LTD., Miyazaki, 883-0062, Japan.
Surg Endosc. 2018 May;32(5):2397-2401. doi: 10.1007/s00464-017-5938-6. Epub 2017 Nov 3.
Iatrogenic inferior vena cava (IVC) injury is a rare but potentially life-threatening complication during laparoscopic surgery. This experimental study aimed to assess the hemostatic ability of a new device, double balloon-equipped central venous (DB-CV) catheter, for IVC injury.
The DB-CV catheter comprises a triple-lumen sphincterotome combined with two dilating balloons having a diameter of 25 mm. The experimental procedures were performed in five pigs. The DB-CV catheter was inserted via the right femoral vein. For the IVC occlusion test, correct placement of the balloons was confirmed by indocyanine green fluorescence imaging, and hemodynamic data were recorded. For the IVC injury test, a 3- to 4-mm circumferential incision was created in IVC, and hemostasis was initiated using balloon inflation 5 s after the injury.
Hemodynamic changes were minimal, with a 20 mmHg reduction in the mean arterial pressure because of IVC occlusion. All bleeding from IVC injuries was successfully temporarily stopped by direct balloon compression, with a mean time to hemostasis of 69 s and mean blood loss of 32 ml. Subsequently, the positioning of IVC injuries between two balloons made it possible to suture the injured IVC.
Balloon occlusion using the DB-CV catheter provides a rapid temporal hemostatic effect and can overcome the serious condition of massive hemorrhage from IVC injuries.
医源性下腔静脉(IVC)损伤是腹腔镜手术中一种罕见但潜在危及生命的并发症。本实验研究旨在评估新型装置——双球囊中央静脉(DB-CV)导管在 IVC 损伤中的止血能力。
DB-CV 导管由三腔括约肌切开器和两个直径为 25mm 的扩张球囊组成。在五只猪中进行了实验程序。DB-CV 导管经右股静脉插入。对于 IVC 闭塞测试,通过吲哚菁绿荧光成像确认球囊的正确位置,并记录血流动力学数据。对于 IVC 损伤测试,在 IVC 上创建一个 3-4mm 周长的切口,在损伤后 5s 开始通过球囊充气进行止血。
由于 IVC 闭塞,平均动脉压仅降低 20mmHg,血流动力学变化很小。通过直接球囊压迫,所有 IVC 损伤的出血均成功暂时停止,止血平均时间为 69s,平均失血量为 32ml。随后,两个球囊之间的 IVC 损伤定位使得缝合损伤的 IVC 成为可能。
使用 DB-CV 导管进行球囊闭塞可提供快速的暂时止血效果,并可克服 IVC 损伤大量出血的严重情况。