Motohashi Yoshikazu, Shimada Ryo, Sasaki Tomoyasu, Katsumata Takahiro, Dan Kazunori, Tsutsui Yasuhiro, Nemoto Shintaro
Department of Thoracic and Cardiovascular Surgery, Osaka Medical College, 2-7 Daigaku-machi, Takatsuki, Osaka, 569-8686, Japan.
Tokai Medical Products Inc., 1485 Sarayashiki-Dengaku, Kasugai, Aichi, 486-0808, Japan.
Gen Thorac Cardiovasc Surg. 2018 Mar;66(3):145-149. doi: 10.1007/s11748-017-0860-3. Epub 2017 Nov 9.
The Blalock-Taussig shunt (BTS) operation is a cornerstone as initial palliative surgery for congenital heart disease with severely reduced pulmonary blood flow (PBF). The ideal PBF provided by BTS is crucial for an uneventful postoperative course, since excess PBF results in acute distress of the systemic circulation and insufficient PBF requires another BTS surgery. Therefore, the goal of this study was to develop a simple device to control the shunt graft flow percutaneously using a constrictor balloon connected to a subcutaneous port.
The device consists of a cylindrical balloon and an anti-bending structure extension connected to the balloon center. A PTFE vascular graft wrapped by the device was connected to a simulated closed circuit to measure the relationship between pressure and blood flow while changing the inner volume of the balloon. In a beagle model of replacement of the right carotid artery, blood flow velocity was measured in the carotid artery after saline injection into the balloon. The blood flow velocity before and after balloon inflation was compared immediately after implantation of the device and at 3 months after implantation.
The device provided good flow control by inflating and deflating the balloon ex vivo and in vivo for up to 3 months in a canine model with a small graft wrapped with the device.
The simple device developed in this study may enable regulation of PBF through a small vascular graft and help to prevent severe morbidity and mortality in the clinical setting of BTS.
布莱洛克 - 陶西格分流术(BTS)手术是先天性心脏病肺血流量严重减少(PBF)时初始姑息性手术的基石。BTS提供的理想肺血流量对于术后平稳恢复至关重要,因为肺血流量过多会导致体循环急性窘迫,而肺血流量不足则需要再次进行BTS手术。因此,本研究的目的是开发一种简单的装置,通过连接皮下端口的收缩球囊经皮控制分流移植物的血流。
该装置由圆柱形球囊和连接到球囊中心的抗弯曲结构延伸部分组成。将包裹有该装置的聚四氟乙烯血管移植物连接到模拟闭合回路,在改变球囊内部容积时测量压力与血流之间的关系。在替换右颈动脉的比格犬模型中,向球囊内注射盐水后测量颈动脉中的血流速度。在植入装置后立即以及植入后3个月比较球囊充气前后的血流速度。
在一个用该装置包裹小移植物的犬模型中,该装置通过在体外和体内对球囊进行充气和放气,在长达3个月的时间内实现了良好的血流控制。
本研究中开发的简单装置可能能够通过小血管移植物调节肺血流量,并有助于在BTS临床环境中预防严重的发病率和死亡率。