Phys Med. 2017 Oct;42:197-202. doi: 10.1016/j.ejmp.2017.09.135. Epub 2017 Oct 7.
To evaluate the dosimetric impact of uncorrected rotations on the planning target volume (PTV) coverage for early stage non-small cell lung cancer patients treated with stereotactic body radiotherapy using Brainlab ExacTrac image guidance.
Twenty-two patients were retrospectively selected. Two scenarios of uncorrected rotations were simulated with magnitude of 1°, 2°, 3° and 5°: (1) rotation around the treatment isocenter; and (2) roll and yaw rotations around a setup isocenter. The D of PTV from recalculated dose on the rotated CT was compared to that from the clinical plan. A logistic regression model was used to predict the probability of dose differences between recalculated and original plans that are less than 2% based on the rotation angle, PTV volume, and distance between the treatment and setup isocenter.
Logistic regression model showed the uncorrected isocentric rotations of up to 2.5° in all directions have negligible dosimetric impact. For non-isocentric rotations, a rotational error of 2° may cause significant under-dose of the PTV. Statistically significant (p<0.05) parameters in the logistic regression model were angle for isocentric rotations, angle and distance for non-isocentric roll rotations, and angle, distance and the PTV volume for non-isocentric yaw rotations.
The severity of the dose deviations due to uncorrected rotations depends on the type and magnitude of the rotation, the volume of the PTV, and the distance between the treatment and setup isocenter, which should be taken into consideration when making clinical judgment of whether the rotational error could be ignored.
评估在使用 Brainlab ExacTrac 图像引导的立体定向体放射治疗早期非小细胞肺癌患者时,未校正旋转对计划靶区(PTV)覆盖的剂量学影响。
回顾性选择了 22 名患者。模拟了未校正旋转的两种情况,旋转幅度为 1°、2°、3°和 5°:(1)绕治疗等中心旋转;(2)绕设置等中心的滚转和偏航旋转。比较了在旋转 CT 上重新计算的剂量与临床计划中 PTV 的 D 值。使用逻辑回归模型,根据旋转角度、PTV 体积和治疗与设置等中心之间的距离,预测重新计算和原始计划之间剂量差异小于 2%的概率。
逻辑回归模型表明,在所有方向上未校正的等中心旋转高达 2.5°,其剂量学影响可以忽略不计。对于非等中心旋转,2°的旋转误差可能导致 PTV 明显欠剂量。逻辑回归模型中的统计学显著(p<0.05)参数为等中心旋转的角度、非等中心滚转旋转的角度和距离,以及非等中心偏航旋转的角度、距离和 PTV 体积。
未校正旋转引起的剂量偏差的严重程度取决于旋转的类型和幅度、PTV 的体积以及治疗与设置等中心之间的距离,在进行旋转误差是否可以忽略的临床判断时,应考虑这些因素。