Suppr超能文献

椎体置换术后相邻节段的前方骨水泥强化可提高环加载下椎体切除 cage 的稳定性:一项生物力学研究。

Anterior cement augmentation of adjacent levels after vertebral body replacement leads to superior stability of the corpectomy cage under cyclic loading-a biomechanical investigation.

机构信息

Center for Orthopaedics and Trauma Surgery, Philipps University Marburg, Germany.

Center for Orthopaedics and Trauma Surgery, Philipps University Marburg, Germany.

出版信息

Spine J. 2018 Mar;18(3):525-531. doi: 10.1016/j.spinee.2017.10.068. Epub 2017 Nov 21.

Abstract

BACKGROUND

In the operative treatment of osteoporotic vertebral body fractures, a dorsal stabilization in combination with a corpectomy of the fractured vertebral body might be necessary with respect to the fracture morphology, whereby the osteoporotic bone quality may possibly increase the risk of implant failure. To achieve better stability, it is recommended to use cement-augmented screws for dorsal instrumentation. Besides careful end plate preparation, cement augmentation of the adjacent end plates has also been reported to lead to less reduction loss.

PURPOSE

The aim of the study was to evaluate biomechanically under cyclic loading whether an additional cement augmentation of the adjacent end plates leads to improved stability of the inserted cage.

STUDY DESIGN/SETTING: Methodical cadaver study.

MATERIALS AND METHODS

Fourteen fresh frozen human thoracic spines with proven osteoporosis were used (T2-T7). After removal of the soft tissues, the spine was embedded in Technovit (Kulzer, Germany). Subsequently, a corpectomy of T5 was performed, leaving the dorsal ligamentary structures intact. After randomization with respect to bone quality, two groups were generated: Dorsal instrumentation (cemented pedicle screws, Medtronic, Minneapolis, MN, USA)+cage implantation (CAPRI Corpectomy Cage, K2M, Leesburg, VA, USA) without additional cementation of the adjacent endplates (Group A) and dorsal instrumentation+cage implantation with additional cement augmentation of the adjacent end plates (Group B). The subsequent axial and cyclic loading was performed at a frequency of 1 Hz, starting at 400 N and increasing the load within 200 N after every 500 cycles up to a maximum of 2,200 N. Load failure was determined when the cages sintered macroscopically into the end plates (implant failure) or when the maximum load was reached.

RESULTS

One specimen in Group B could not be clamped appropriately into the test bench for axial loading because of a pronounced scoliotic misalignment and had to be excluded. The mean strength for implant failure was 1,000 N±258.2 N in Group A (no cement augmentation of the adjacent end plates, n=7); on average, 1,622.1±637.6 cycles were achieved. In Group B (cement augmentation of the adjacent end plates, n=6), the mean force at the end of loading was 1,766.7 N±320.4 N; an average of 3,572±920.6 cycles was achieved. Three specimens reached a load of 2,000 N. The differences between the two groups were significant (p=.006 and p=.0047) regarding load failure and number of cycles.

CONCLUSIONS

Additional cement augmentation of the adjacent end plates during implantation of a vertebral body replacement in osteoporotic bone resulted in a significant increased stability of the cage in the axial cyclic loading test.

摘要

背景

在骨质疏松性椎体骨折的手术治疗中,根据骨折形态,可能需要进行背侧稳定化治疗,结合骨折椎体的切除术,而骨质疏松的骨质量可能会增加植入物失败的风险。为了获得更好的稳定性,建议使用骨水泥增强螺钉进行背侧器械固定。除了仔细准备终板外,已有报道称,相邻终板的骨水泥增强也可导致更少的复位丢失。

目的

本研究旨在通过循环加载下的生物力学评估,确定相邻终板的额外骨水泥增强是否可提高植入笼的稳定性。

研究设计/设置:方法学尸体研究。

材料和方法

使用了 14 个经证实患有骨质疏松症的新鲜冷冻人胸段脊柱(T2-T7)。去除软组织后,脊柱被嵌入 Technovit(德国库尔泽)。随后,进行 T5 的椎体切除术,保留背侧韧带结构完整。随机分组后,根据骨质量生成了两组:背侧器械固定(骨水泥增强椎弓根螺钉,美敦力,明尼苏达州明尼阿波利斯)+植入笼(CAPRI 椎体切除术笼,K2M,弗吉尼亚州利斯堡),不进行相邻终板的额外骨水泥增强(A 组)和背侧器械固定+相邻终板的额外骨水泥增强(B 组)。随后以 1 Hz 的频率进行轴向和循环加载,起始负荷为 400 N,并在每 500 次循环后增加 200 N 的负荷,直到最大负荷达到 2200 N。当笼烧结宏观上与终板融合(植入物失效)或达到最大负荷时,确定负载失效。

结果

B 组中有 1 个标本因明显的脊柱侧凸错位而无法适当固定在试验台上,因此必须排除在外。A 组(不进行相邻终板的骨水泥增强,n=7)的植入物失效平均强度为 1000 N±258.2 N;平均达到 1622.1±637.6 个循环。B 组(相邻终板的骨水泥增强,n=6)的加载结束时的平均力为 1766.7 N±320.4 N;平均达到 3572±920.6 个循环。3 个标本达到 2000 N 的负荷。两组之间在负载失效和循环次数方面差异具有统计学意义(p=.006 和 p=.0047)。

结论

在骨质疏松性骨中植入椎体置换物时,相邻终板的额外骨水泥增强可显著提高笼在轴向循环加载试验中的稳定性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验