Suppr超能文献

纵向位置对心脏电影平衡稳态自由进动成像中稳态信号演变的影响。

Influence of longitudinal position on the evolution of steady-state signal in cardiac cine balanced steady-state free precession imaging.

作者信息

Spear Tyler J, Stromp Tori A, Leung Steve W, Vandsburger Moriel H

机构信息

Saha Cardiovascular Research Center, University of Kentucky, Lexington, KY, USA.

Department of Physiology, University of Kentucky, Lexington, KY, USA.

出版信息

Acta Radiol Open. 2017 Nov 21;6(11):2058460117729186. doi: 10.1177/2058460117729186. eCollection 2017 Nov.

Abstract

BACKGROUND

Emerging quantitative cardiac magnetic resonance imaging (CMRI) techniques use cine balanced steady-state free precession (bSSFP) to measure myocardial signal intensity and probe underlying physiological parameters. This correlation assumes that steady-state is maintained uniformly throughout the heart in space and time.

PURPOSE

To determine the effects of longitudinal cardiac motion and initial slice position on signal deviation in cine bSSFP imaging by comparing two-dimensional (2D) and three-dimensional (3D) acquisitions.

MATERIAL AND METHODS

Nine healthy volunteers completed cardiac MRI on a 1.5-T scanner. Short axis images were taken at six slice locations using both 2D and 3D cine bSSFP. 3D acquisitions spanned two slices above and below selected slice locations. Changes in myocardial signal intensity were measured across the cardiac cycle and compared to longitudinal shortening.

RESULTS

For 2D cine bSSFP, 46% ± 9% of all frames and 84% ± 13% of end-diastolic frames remained within 10% of initial signal intensity. For 3D cine bSSFP the proportions increased to 87% ± 8% and 97% ± 5%. There was no correlation between longitudinal shortening and peak changes in myocardial signal. The initial slice position significantly impacted peak changes in signal intensity for 2D sequences ( < 0.001).

CONCLUSION

The initial longitudinal slice location significantly impacts the magnitude of deviation from steady-state in 2D cine bSSFP that is only restored at the center of a 3D excitation volume. During diastole, a transient steady-state is established similar to that achieved with 3D cine bSSFP regardless of slice location.

摘要

背景

新兴的定量心脏磁共振成像(CMRI)技术利用电影平衡稳态自由进动(bSSFP)来测量心肌信号强度并探究潜在的生理参数。这种相关性假定稳态在整个心脏的空间和时间上均一维持。

目的

通过比较二维(2D)和三维(3D)采集,确定心脏纵向运动和初始切片位置对电影bSSFP成像中信号偏差的影响。

材料与方法

9名健康志愿者在1.5-T扫描仪上完成心脏MRI检查。使用2D和3D电影bSSFP在6个切片位置采集短轴图像。3D采集跨越选定切片位置上方和下方的两个切片。在心动周期中测量心肌信号强度的变化,并与纵向缩短进行比较。

结果

对于2D电影bSSFP,所有帧的46%±9%和舒张末期帧的84%±13%保持在初始信号强度的10%以内。对于3D电影bSSFP,这一比例分别增至87%±8%和97%±5%。纵向缩短与心肌信号的峰值变化之间无相关性。初始切片位置对2D序列的信号强度峰值变化有显著影响(<0.001)。

结论

初始纵向切片位置显著影响2D电影bSSFP中偏离稳态的幅度,这种偏差仅在3D激发容积的中心恢复。在舒张期,无论切片位置如何,都会建立一个类似于3D电影bSSFP所实现的瞬态稳态。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d7a7/5700791/2cb935d9e8bb/10.1177_2058460117729186-fig1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验