常用于确定图像质量的乳腺造影体模:一项比较研究。

Mammographic Phantoms Frequently Used to Determine Image Quality: A Comparative Study.

作者信息

AlKhalifah Khaled, Brindabhan Ajit

机构信息

Radiologic Sciences Dep., Kuwait University, Faculty of Allied Health Sciences, PO Box 31470, Sulaibikhat 90805, Kuwait. Tel +965-94495959.

出版信息

J Allied Health. 2017 Winter;46(4):239-242.

Abstract

UNLABELLED

Appropriate diagnosis depends on the image quality of the mammographic system. Phantoms are normally used, since they have physical properties similar to those of different human tissues, and are important tools for quality assurance of mammography equipment, evaluation of image quality, and accurate determination of patient dosage. The aim of this study was to evaluate various mammographic phantoms used to determine the image quality of digital mammographic equipment.

METHODS

Three phantoms were used: an American College of Radiology accreditation phantom, an MTM 100/R, and a tissue-equivalent phantom. Tungsten/silver target/filter material was used with four variable exposures (25, 28, 30, and 32 kVp). In total, there were 12 images. The images were evaluated by scoring the visibility of the structures, and a total score out of 100% was recorded. Statistical analysis was conducted using non-parametric tests.

RESULTS

There was no significant difference in the visibility of structures among the phantoms at different kVp. However, there were significant differences (p=0.04) in the visibility of the tumor masses when the kVp was varied. In addition, visibility rates differed significantly between the three phantoms (p≤0.05).

CONCLUSION

All the phantoms used in this study were suitable for evaluating digital mammography systems. However, the exposure setting for evaluation must be carefully considered. The tissue-equivalent phantom performed better in terms of the visibility rate to assess image quality in the digital mammography units.

摘要

未标注

准确的诊断取决于乳腺摄影系统的图像质量。通常使用体模,因为它们具有与不同人体组织相似的物理特性,是乳腺摄影设备质量保证、图像质量评估以及准确确定患者剂量的重要工具。本研究的目的是评估用于确定数字乳腺摄影设备图像质量的各种乳腺摄影体模。

方法

使用了三种体模:美国放射学会认证体模、MTM 100/R和组织等效体模。使用钨/银靶/滤过材料并设置四种可变曝光量(25、28、30和32 kVp)。总共获得12幅图像。通过对结构的可见性进行评分来评估图像,并记录100%的总分。使用非参数检验进行统计分析。

结果

在不同kVp下,各体模之间结构的可见性无显著差异。然而,当kVp变化时,肿瘤块的可见性存在显著差异(p = 0.04)。此外,三种体模之间的可见率差异显著(p≤0.05)。

结论

本研究中使用的所有体模都适用于评估数字乳腺摄影系统。然而,必须仔细考虑评估的曝光设置。在评估数字乳腺摄影设备的图像质量时,组织等效体模在可见率方面表现更好。