Saito H, Saka H, Sakai S, Shimokata K
Department of Medicine, National Chubu Hospital, Aichi, Japan.
Chest. 1989 Mar;95(3):700-1. doi: 10.1378/chest.95.3.700.
Transbronchial lung biopsy performed on a 51-year-old woman for the evaluation of nodular densities with cavities in the bilateral apical areas was complicated by the breakage of the biopsy forceps. The broken fragment of the forceps was successfully retrieved by use of the magnetic extractor through a fiberoptic bronchoscope under fluoroscopic guidance.
对一名51岁女性进行经支气管肺活检,以评估双侧肺尖区域有空洞的结节状密度影,活检过程中活检钳发生断裂。在荧光透视引导下,通过纤维支气管镜使用磁性取出器成功取出了活检钳的断裂碎片。