Department of Radiation Oncology, Rush University Medical Center, Chicago, Illinois 60612, United States of America.
Phys Med Biol. 2018 Jan 11;63(2):025016. doi: 10.1088/1361-6560/aaa200.
By collimating the photons scattered when a megavoltage therapy beam interacts with the patient, a Compton-scatter image may be formed without the delivery of an extra dose. To characterize and assess the potential of the technique, an analytical model for simulating scatter images was developed and validated against Monte Carlo (MC). For three phantoms, the scatter images collected during irradiation with a 6 MV flattening-filter-free therapy beam were simulated. Images, profiles, and spectra were compared for different phantoms and different irradiation angles. The proposed analytical method simulates accurate scatter images up to 1000 times faster than MC. Minor differences between MC and analytical simulated images are attributed to limitations in the isotropic superposition/convolution algorithm used to analytically model multiple-order scattering. For a detector placed at 90° relative to the treatment beam, the simulated scattered photon energy spectrum peaks at 140-220 keV, and 40-50% of the photons are the result of multiple scattering. The high energy photons originate at the beam entrance. Increasing the angle between source and detector increases the average energy of the collected photons and decreases the relative contribution of multiple scattered photons. Multiple scattered photons cause blurring in the image. For an ideal 5 mm diameter pinhole collimator placed 18.5 cm from the isocenter, 10 cGy of deposited dose (2 Hz imaging rate for 1200 MU min treatment delivery) is expected to generate an average 1000 photons per mm at the detector. For the considered lung tumor CT phantom, the contrast is high enough to clearly identify the lung tumor in the scatter image. Increasing the treatment beam size perpendicular to the detector plane decreases the contrast, although the scatter subject contrast is expected to be greater than the megavoltage transmission image contrast. With the analytical method, real-time tumor tracking may be possible through comparison of simulated and acquired patient images.
通过准直兆伏级治疗射线与患者相互作用时散射的光子,可以在不额外增加剂量的情况下形成康普顿散射图像。为了对该技术进行特征描述和评估,开发了一种用于模拟散射图像的分析模型,并通过蒙特卡罗(MC)进行了验证。对于三个体模,模拟了用 6 MV 无均整滤过治疗射线照射时采集的散射图像。比较了不同体模和不同照射角度的图像、轮廓和光谱。与 MC 相比,所提出的分析方法可以快 1000 倍模拟准确的散射图像。MC 和分析模拟图像之间的微小差异归因于用于分析模拟多次散射的各向同性叠加/卷积算法的局限性。对于放置在相对于治疗束 90°的探测器,模拟的散射光子能谱在 140-220keV 处峰值,并且 40-50%的光子是多次散射的结果。高能光子起源于束入口处。增加源和探测器之间的角度会增加采集光子的平均能量,并降低多次散射光子的相对贡献。多次散射光子会导致图像模糊。对于放置在离等中心 18.5cm 处的理想 5mm 直径针孔准直器,预计 10cGy 的沉积剂量(1200MU min 治疗输送的 2Hz 成像速率)将在探测器上产生平均每毫米 1000 个光子。对于考虑的肺部肿瘤 CT 体模,散射图像中可以清晰地识别出肺部肿瘤,对比度足够高。增加垂直于探测器平面的治疗束尺寸会降低对比度,尽管散射对象对比度预计会大于兆伏级透射图像对比度。通过分析方法,通过比较模拟和获取的患者图像,可能实现实时肿瘤跟踪。