Fracture of a fenestrated metal backing of a tibial knee component. A case report.

作者信息

Gradisar I A, Hoffmann M L, Askew M J

机构信息

Department of Orthopaedic Surgery, Akron City Hospital, Ohio 44309.

出版信息

J Arthroplasty. 1989;4(1):27-30. doi: 10.1016/s0883-5403(89)80050-6.

Abstract

The case of a metal-backed tibial component that failed by metal fracture is reported. Fracture occurred through fenestrations in the metal plate surrounding the central metal stem. The fracture followed loss of bony support beneath the medial tibial plateau, which had allowed varus deformity of the replaced joint. A high tibial osteotomy procedure had preceded replacement of the joint.