Abstract

INTRODUCTION

The goal of this study was to examine the potential value of corneal lubrication in improving the quality of optical coherence tomography (OCT) imaging of the retina.

MATERIALS AND METHODS

This study is a consecutive series of 13 eyes of 11 patients in whom repeated attempts at OCT imaging failed to yield a good quality study despite the absence of significant media opacity or inadequate pupil dilation. Immediately following several poor quality scans, each eye received lubricating eye drops. The quality of images before and after the administration of drops was assessed.

RESULTS

A statistically significant improvement in OCT image quality was observed following the administration of eye drops in each case. The change in mean signal strength (SS) was from 4.35 to 6.26 (p=0.0002). The proportion of scans with erroneous edge detection decreased from 41 to 13 percent.

DISCUSSION

Lubricating drops appear to improve the quality and feasibility of OCT imaging in selected cases.

摘要

引言

本研究的目的是探讨角膜润滑在提高视网膜光学相干断层扫描(OCT)成像质量方面的潜在价值。

材料与方法

本研究为连续纳入11例患者的13只眼,尽管不存在明显的介质混浊或瞳孔散大不足,但多次尝试进行OCT成像均未能获得高质量的研究结果。在几次质量较差的扫描后,立即给每只眼滴入润滑眼药水。评估滴眼药水前后图像的质量。

结果

在每种情况下,滴入眼药水后OCT图像质量均有统计学意义的改善。平均信号强度(SS)的变化从4.35变为6.26(p = 0.0002)。边缘检测错误的扫描比例从41%降至13%。

讨论

在特定情况下润滑眼药水似乎可提高OCT成像的质量和可行性。

相似文献

1
The effect of lubricating eye drops on optical coherence tomography imaging of the retina.
Digit J Ophthalmol. 2009 May 30;15(2):24-26. doi: 10.5693/djo.01.2009.006. eCollection 2009.
3
Spectral-domain optical coherence tomography with multiple B-scan averaging for enhanced imaging of retinal diseases.
Ophthalmology. 2008 Jun;115(6):1071-1078.e7. doi: 10.1016/j.ophtha.2007.09.001. Epub 2007 Dec 3.
4
Effect of contact lens on optical coherence tomography imaging of rodent retina.
Curr Eye Res. 2013 Dec;38(12):1235-40. doi: 10.3109/02713683.2013.815218. Epub 2013 Sep 3.
6
A model for the effect of disturbances in the optical media on the OCT image quality.
Invest Ophthalmol Vis Sci. 2009 Feb;50(2):787-92. doi: 10.1167/iovs.08-2364. Epub 2008 Sep 4.
7
Correction of ocular shape in retinal optical coherence tomography and effect on current clinical measures.
Am J Ophthalmol. 2013 Aug;156(2):304-11. doi: 10.1016/j.ajo.2013.03.012. Epub 2013 May 6.
8
Optical coherence tomography (OCT) for detection of macular oedema in patients with diabetic retinopathy.
Cochrane Database Syst Rev. 2011 Jul 6(7):CD008081. doi: 10.1002/14651858.CD008081.pub2.
10
Effect of posterior capsular opacification removal on macular optical coherence tomography.
Eur J Ophthalmol. 2008 May-Jun;18(3):435-41. doi: 10.1177/112067210801800319.

引用本文的文献

1
A Deep Learning Approach to Denoise Optical Coherence Tomography Images of the Optic Nerve Head.
Sci Rep. 2019 Oct 8;9(1):14454. doi: 10.1038/s41598-019-51062-7.

本文引用的文献

1
Effect of corneal drying on optical coherence tomography.
Ophthalmology. 2006 Jun;113(6):985-991. doi: 10.1016/j.ophtha.2006.02.018.
2
A new quality assessment parameter for optical coherence tomography.
Br J Ophthalmol. 2006 Feb;90(2):186-90. doi: 10.1136/bjo.2004.059824.
3
Errors in retinal thickness measurements obtained by optical coherence tomography.
Ophthalmology. 2006 Feb;113(2):285-93. doi: 10.1016/j.ophtha.2005.10.005. Epub 2006 Jan 10.
4
Repeatability and reproducibility of fast macular thickness mapping with stratus optical coherence tomography.
Arch Ophthalmol. 2005 Oct;123(10):1330-7. doi: 10.1001/archopht.123.10.1330.
5
Reproducibility of optic nerve head measurements obtained by optical coherence tomography.
Eur J Ophthalmol. 2005 Jul-Aug;15(4):486-92. doi: 10.1177/112067210501500412.
6
Evaluation of image artifact produced by optical coherence tomography of retinal pathology.
Am J Ophthalmol. 2005 Jan;139(1):18-29. doi: 10.1016/j.ajo.2004.07.050.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验