Karelitz Joshua L, Michael Valerie C, Perkins Kenneth A
Department of Psychiatry of the University of Pittsburgh, Pittsburgh, Pennsylvania.
J Smok Cessat. 2017 Jun;12(2):105-112. doi: 10.1017/jsc.2015.18. Epub 2016 Feb 2.
The current study examined the level of agreement in expired-air carbon monoxide (CO) values, focusing especially on those confirming abstinence, between the two most commonly used CO monitors, the Vitalograph BreathCO and the Bedfont piCO+ Smokerlyzer.
Expired-air samples were collected via both monitors from adult dependent smokers (44 M, 34 F) participating in studies using CO values to confirm abstinence durations of: 24 hours, 12 hours, or no abstinence. All met DSM-IV nicotine dependence criteria and had a mean (SD) Fagerström Test of Cigarette Dependence score of 5.1 (1.8). Paired data collected across multiple visits were analyzed by regression-based Bland-Altman method of Limits of Agreement.
Analysis indicated a lack of agreement in CO measurement between monitors. Overall, the Bedfont monitor gave mean (±SEM) readings 3.83 (±.23) ppm higher than the Vitalograph monitor. Mean differences between monitors were larger for those ad lib smoking (5.65±.38 ppm) than those abstaining 12-24 hours (1.71±.13 ppm). Yet, there also was not consistent agreement in classification of 24 hour abstinence between monitors.
Systematic differences in CO readings demonstrate these two very common monitors may not result in interchangeable values, and reported outcomes in smoking research based on CO values may depend on the monitor used.
本研究检测了两种最常用的一氧化碳(CO)监测仪,即维他福呼吸一氧化碳监测仪(Vitalograph BreathCO)和贝德丰piCO + 吸烟检测分析仪(Bedfont piCO+ Smokerlyzer)对呼出气中CO值的一致性水平,尤其关注那些确认戒烟情况的CO值。
通过这两种监测仪收集成年依赖吸烟者(44名男性,34名女性)的呼出气样本,这些吸烟者参与了使用CO值来确认戒烟时长的研究,戒烟时长分别为24小时、12小时或未戒烟。所有参与者均符合《精神疾病诊断与统计手册》第四版(DSM-IV)尼古丁依赖标准,平均(标准差)法格斯特罗姆烟草依赖测试(Fagerström Test of Cigarette Dependence)得分为5.1(1.8)。通过基于回归的一致性界限布兰德-奥特曼方法(Bland-Altman method of Limits of Agreement)分析多次访视收集的配对数据。
分析表明,两种监测仪在CO测量方面缺乏一致性。总体而言,贝德丰监测仪给出的平均(±标准误)读数比维他福监测仪高3.83(±0.23)ppm。对于那些随意吸烟的人,监测仪之间的平均差异(5.65±0.38 ppm)大于那些戒烟12 - 24小时的人(1.71±0.13 ppm)。然而在监测仪之间,对于24小时戒烟的分类也没有一致的结果。
CO读数的系统差异表明,这两种非常常见的监测仪可能不会得出可互换的值,并且基于CO值的吸烟研究报告结果可能取决于所使用的监测仪。