Trippestad A
Acta Chir Scand. 1985;151(4):391-5.
Late fibre breakage with longitudinal tears occurred in 4 patients of a series of some 300 implanted knitted Dacron double velour arterial prostheses. The four patients all received their grafts for occlusive aortoiliac disease in the year 1978, and the rupture was found 4 to 6 years later. Three of the failures occurred in axillo-bifemoral bypass grafts, whereas the fourth patient had a bifurcated aortic prosthesis. Dilatation was significant in the latter graft only. Structural weakness might appear during production, during later handling, or possibly as a result of interference from infected material or host factors. The cause of graft failures in the present series could not be ascertained. Whatever type of knitted Dacron used, a 2-3% incidence of graft deterioration might be expected. Therefore regular, life-long follow-up of patients with Dacron arterial prostheses seems warranted.
在一系列约300例植入的针织涤纶双绒动脉假体中,有4例出现了晚期纤维断裂并伴有纵向撕裂。这4例患者均于1978年因闭塞性主髂动脉疾病接受了移植手术,4至6年后发现破裂。其中3例失败发生在腋-双股旁路移植术中,而第4例患者使用的是分叉型主动脉假体,仅后者的移植物出现了明显扩张。结构薄弱可能在生产过程中、后期处理过程中出现,也可能是由于感染物质或宿主因素的干扰所致。本系列中移植物失败的原因无法确定。无论使用何种类型的针织涤纶,预计移植物恶化的发生率为2%-3%。因此,对使用涤纶动脉假体的患者进行定期、终身随访似乎是必要的。