经三维经食管超声心动图评估的Memo3D二尖瓣成形环的瓣环动力学

Annular dynamics of memo3D annuloplasty ring evaluated by 3D transesophageal echocardiography.

作者信息

Nishi Hiroyuki, Toda Koichi, Miyagawa Shigeru, Yoshikawa Yasushi, Fukushima Satsuki, Yoshioka Daisuke, Sawa Yoshiki

机构信息

Department of Cardiovascular Surgery, Osaka University Graduate School of Medicine, 2-2 Yamada-oka, Suita, Osaka, 565-0871, Japan.

出版信息

Gen Thorac Cardiovasc Surg. 2018 Apr;66(4):214-219. doi: 10.1007/s11748-018-0886-1. Epub 2018 Jan 17.

Abstract

BACKGROUND

We assessed the mitral annular motion after mitral valve repair with the Sorin Memo 3D® (Sorin Group Italia S.r.L., Saluggia, Italy), which is a unique complete semirigid annuloplasty ring intended to restore the systolic profile of the mitral annulus while adapting to the physiologic dynamism of the annulus, using transesophageal real-time three-dimensional echocardiography.

METHODS

17 patients (12 male; mean age 60.4 ± 14.9 years) who underwent mitral annuloplasty using the Memo 3D ring were investigated. Mitral annular motion was assessed using QLAB®version8 allowing for a full evaluation of the mitral annulus dynamics. The mitral annular dimensions were measured throughout the cardiac cycle using 4D MV assessment2® while saddle shape was assessed through sequential measurements by RealView®.

RESULTS

Saddle shape configuration of the mitral annulus and posterior and anterior leaflet motion could be observed during systole and diastole. The mitral annular area changed during the cardiac cycle by 5.7 ± 1.8%.The circumference length and diameter also changed throughout the cardiac cycle. The annular height was significantly higher in mid-systole than in mid-diastole (p < 0.05).

CONCLUSIONS

The Memo 3D ring maintained a physiological saddle-shape configuration throughout the cardiac cycle. Real-time three-dimensional echocardiography analysis confirmed the motion and flexibility of the Memo 3D ring upon implantation.

摘要

背景

我们使用经食管实时三维超声心动图,评估了采用索林Memo 3D®(索林集团意大利有限公司,意大利萨卢贾)进行二尖瓣修复术后的二尖瓣环运动情况。索林Memo 3D®是一种独特的完整半刚性瓣环成形环,旨在恢复二尖瓣环的收缩形态,同时适应瓣环的生理动态变化。

方法

对17例使用Memo 3D环进行二尖瓣环成形术的患者(12例男性;平均年龄60.4±14.9岁)进行了研究。使用QLAB®version8评估二尖瓣环运动,以全面评估二尖瓣环动力学。在整个心动周期中使用4D MV assessment2®测量二尖瓣环尺寸,同时通过RealView®进行连续测量来评估鞍形。

结果

在收缩期和舒张期均可观察到二尖瓣环的鞍形结构以及前后叶运动。二尖瓣环面积在心动周期中变化了5.7±1.8%。周长和直径在整个心动周期中也发生了变化。收缩中期的瓣环高度显著高于舒张中期(p<0.05)。

结论

Memo 3D环在整个心动周期中保持了生理鞍形结构。实时三维超声心动图分析证实了Memo 3D环植入后的运动和灵活性。