Magn Reson Med. 2018 Sep;80(3):1125-1131. doi: 10.1002/mrm.27080. Epub 2018 Jan 25.
Cerebral venous oxygenation (Y) is an important physiological parameter and has potential clinical application in many brain diseases. T-relaxation-under-spin-tagging (TRUST) is a commonly used MRI method to measure Y. Harmonization of this technique across MRI vendors is important for dissemination and multi-center studies of brain oxygenation and metabolism as a disease biomarker.
TRUST pulse sequence components and imaging parameters were carefully matched between two major MRI vendors, Philips and Siemens. Each subject (=10) was scanned on both scanners within a 2.5-hour period. On each scanner, the subject was scanned in two sessions to assess inter-session reproducibility. A hyperoxia challenge was also included in both sessions and on both scanners to evaluate the sensitivity of the technique to Y changes. Measured Y values, confidence interval of Y estimates (), as well as intra-session and inter-session coefficient of variation (CoV) of Y, were compared between the two scanners.
Y measured on the two vendors were highly compatible and strongly correlated (=0.957). Y changes associated with hyperoxia challenge were significant on both scanners (<0.001) and were also correlated across scanners (=0.007). Intra-session and inter-session CoV of measured Y were less than 3% and showed no difference between scanners. were less than 1% on both scanners and showed no difference between scanners when echo times were matched on the two scanners.
This work suggests that harmonized TRUST MRI can yield highly compatible Y measurements across different vendors.
脑静脉血氧饱和度(Y)是一个重要的生理参数,在许多脑部疾病中有潜在的临床应用。T 弛豫下自旋标记(TRUST)是一种常用的 MRI 方法,用于测量 Y。这项技术在不同 MRI 供应商之间的协调对于脑氧合和代谢作为疾病生物标志物的传播和多中心研究非常重要。
在飞利浦和西门子这两家主要的 MRI 供应商之间,仔细匹配了 TRUST 脉冲序列组件和成像参数。每个被试(=10)在 2.5 小时内分别在两台扫描仪上进行扫描。在每台扫描仪上,被试在两个会话中进行扫描,以评估会话间的可重复性。还在两个会话和两台扫描仪上进行了高氧挑战,以评估该技术对 Y 变化的敏感性。比较了两台扫描仪之间的 Y 测量值、Y 估计值的置信区间()以及 Y 的会话内和会话间变异系数(CoV)。
两台供应商的 Y 测量值高度兼容且具有很强的相关性(=0.957)。高氧挑战引起的 Y 变化在两台扫描仪上均具有统计学意义(<0.001),并且在扫描仪之间也具有相关性(=0.007)。测量 Y 的会话内和会话间 CoV 小于 3%,且在扫描仪之间无差异。在两台扫描仪上的均小于 1%,当在两台扫描仪上匹配回波时间时,在扫描仪之间也无差异。
这项工作表明,经过协调的 TRUST MRI 可以在不同供应商之间产生高度兼容的 Y 测量值。